Performing A Calibration - Agilent Technologies ENA Series User Manual

Rf network analyzer
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Setting Description
Trace-2 measurement parameter: S
Allocate a trace to upper and lower displays.
Auto-scale all traces.
Figure 14-27
Results of Unbalanced Measurements

3. Performing a Calibration

Perform a full three-port calibration for the three ports to be used.
Step 1. Set the type and conditions of the calibration.
Setting Description
Calibration kit to use: 85033D
Type of calibration: Full three-port calibration
Test ports to calibrate: 1, 2, 3
Step 2. Perform a reflection calibration.
Setting Description
Select reflection calibration.
Perform Port 1 calibration.
Chapter 14
Measuring an Unbalanced and Balanced Bandpass Filter
Keystroke
31
Keystroke
Calibrate
Select Ports
Keystroke
Reflection
(With the OPEN connected)
(With the SHORT connected)
Measurement Examples
-
S31
-
-
Allocate Traces
-
Auto Scale All
-
-
Cal Kit
85033D
-
3-Port Cal
-
(check only)
1-2-3
Port 1 OPEN
Port 1 SHORT
309