For calibration with the E5091A connected, because the setting of the E5091 can be
performed for each channel, you need to change the ports to which the standard is
connected for the channel even for the same measurement parameter.

Operational procedure

Step 1. Press
calibration to the active channel.
Step 2. Follow "Displaying the E5091A property" on page 273 to display the E5091A property.
Step 3. According to Chapter 4, "Calibration," on page 77, perform calibration. Check the
connected test ports shown in the calibration property as the port names of the
E5070A/E5071A on the calibration menu, connect the calibration standard to the
corresponding test ports of the E5091A, and perform calibration.