Step 9. After restart, the Enter Network Password dialog box as shown in Figure 12-21 appears.
Enter nothing and click the
Figure 12-21
Enter Network Password dialog box
NOTE
If you enter the password here, each time you start the instrument, the Enter Network
Password dialog box appears and asks you to enter the password. If you click the
button, the Enter Network Password dialog box will appear again at the next startup.
Step 10. The System Setup Complete dialog box appears. Click the
Note that, if the system calibration data file has a problem, the Recovery Failed dialog box
as shown in Figure 12-22 appears instead of the System Setup Complete dialog box, and
then the System Setup Incomplete dialog box appears.
Figure 12-22
Recovery Failed dialog box
NOTE
If the Recovery Failed dialog box and the System Setup Incomplete dialog box appear or if
the problem persists even if you executing the system recovery, the instrument is at fault.
Contact your local Agilent Technologies customer center listed at the end of this manual or
distributor.
Step 11. For the E5070A/E5071A equipped with the Option 016 touch screen, execute the
calibration of the touch screen. For information on the execution procedure, refer to
"Calibration of the Touch Screen" on page 268.
Now the system recovery of the E5070A/E5071A is complete.
Chapter 12
Setting and Using the Control and Management Functions
button.
OK
System Recovery
Cancel
button to restart.
OK
267