Specifications and Supplemental Information
Corrected System Performance
Table 15-4
Corrected System Performance With 3.5 mm Device Connector Type, 85033E
Calibration Kit
Network analyzer: E5070A/E5071A, Calibration kit: 85033E (3.5 mm, 50 W), Calibration: full 2-port
IF bandwidth = 10 Hz, No averaging applied to data, Environmental temperature = 23°C±5°C with <1°C deviation
from calibration temperature, Isolation calibration not omitted
Description
Directivity
Source Match
Load Match
Reflection Tracking
Transmission Tracking
Transmission Uncertainty (Specification)
Reflection Uncertainty (Specification)
332
3 MHz to 3 GHz
46
43
46
±0.006
±0.016
Specification (dB)
3 GHz to 6 GHz
38
37
38
±0.009
±0.065
6 GHz to 8.5 GHz
38
36
38
±0.010
±0.069