Softkey Functions

Trigger Menu

Trigger Menu
Key Operation
Displays following seven softkeys for setting the trigger. Once the trigger mode is set, measurements are executed
according to the trigger mode even when the channel is no longer on display due to reduction of the number of
channels to be displayed from the "Display Menu" on page 424.
Sets the active channel trigger mode to "hold". A trigger sent from the
Hold
trigger source to that channel will not prompt a sweep.
Sets the active channel trigger mode to "single". A trigger sent from the
Single
trigger source to that channel prompts a single sweep and then the sweep
mode changes to "hold."
Sets the active channel trigger mode to "continuous". Each trigger sent
Continuous
from the trigger source to that channel prompts a single sweep.
Sets all channel trigger modes to "hold".
Hold All Channels
Sets all displayed channel trigger modes to "continuous". See "Number of
Continuous Disp
Channels and Channel Window Arrangement" on page 56 for details about
Channels
displayed channels.
Displays softkeys for selecting the trigger source.
Trigger Source
Selects the "internal" trigger source of the analyzer. The analyzer will
Internal
generate a series of trigger signals. A trigger, once generated, is sent to all
channels in order no matter how many channels are displayed.
Selects the "external" trigger source. A signal input to the external trigger
External
input terminal (BNC(f) connector) on the rear panel will be used as the
trigger source. A trigger, once generated, is sent to all channels in order no
matter how many channels are displayed.
Sets the trigger source to "manual". Pressing the
Manual
Menu generates a trigger. A trigger, once generated, is sent to all channels
in order no matter how many channels are displayed.
Sets the analyzer trigger source to "bus". A trigger is generated by sending
Bus
a trigger command through the GPIB or a LAN. A trigger, once generated,
is sent to all channels in order no matter how many channels are displayed.
Returns to the softkey display screen one level higher.
Cancel
Aborts a sweep.
Restart
When the trigger source is set to "manual", generates a trigger.
Trigger
Returns to the "E5070A/E5071A Menu (Top Menu)" on page 404.
Return
452
Function
key in the Trigger
Trigger
SCPI Command
:INIT{1-9}:CONT
:INIT{1-9}
:INIT{1-9}:CONT
None
None
:TRIG:SOUR INT
:TRIG:SOUR EXT
:TRIG:SOUR MAN
:TRIG:SOUR BUS
:ABOR
:TRIG