Step 15. Select the class to be redefined.
Softkey
Open
Short
Load
Thru
Step 16. Select a test port.
Softkey
Port 1
Port 2
Port 3
Port 4
NOTE
Select
Step 17. Select the standards to be registered in the class from among standards numbered 1 through
21.
Step 18. Repeat Step 16 and Step 17 until classes are defined for all test ports that need to be
redefined.
Step 19. Press
Step 20. Repeat Step 15 to Step 19 to redefine all classes that need to be modified.
Step 21. Press
Step 22. Press
the screen.
Chapter 4
Function
Selects the OPEN class.
Selects the SHORT class.
Selects the LOAD class.
Selects the THRU class.
Function
Selects port 1.
Selects port 2.
Selects port 3.
Selects port 4.
to use the same standards for all test ports.
Set All
.
Return
.
Return
and input a new label for the calibration kit using the keypad displayed on
Label Kit
Changing the Calibration Kit Definition
Calibration
115