Overview of Functions
Screen Area: Names and Functions of Parts
5-5. Trace Status Area
The setup for each trace is displayed here.
Table 2-1
Trace Status Display
Classification
Error correction
Turning on/off traces
Performing data math
Electrical delay
Smoothing
Parameter conversion
5-6. Reference Line Indicators
The indicators that indicate the position of the reference line for the Y-axis scale in the
rectangular display format. One indicator to the right and the other to the left of the scale.
( and
entry bar using keys:
reference line by placing the mouse pointer on either of the two reference line indicators
(the pointer changes from
button kept pressed, and then releasing the button at the desired location. (This mouse
operation is called a drag-and-drop operation).
5-7. Trace Number
In the rectangular display format, the trace number is displayed in the same color as the
trace at the right end of each trace.
42
Contents inside the [ ]
RO
RS
RT
F1
F2
F3
F4
Nothing
M
D&M
off
(
)
D+M
D+M&M
(
)
D-M
D-M&M
(
)
D*M
D*M&M
(
)
D/M
D/M&M
Del
Smo
Zr
Zt
Yr
Yt
1/S
). To enter a numeric value for the position of the reference line, open the data
Meaning
Error correction: ON (OPEN (n) response calibration)
Error correction: ON (SHORT (n) response calibration)
Error correction: ON (THRU (n) response calibration)
Error correction: ON (1-port calibration)
Error correction: ON (Full 2-port calibration)
Error correction: ON (Full 3-port calibration)
Error correction: ON (Full 4-port calibration)
Data trace: ON, Memory trace: OFF
Data trace: OFF, Memory trace: ON
Data trace: ON, Memory trace: ON
Data trace: OFF, Memory trace: OFF
Execution of Data+Mem math
Execution of Data-Mem math
Execution of Data*Mem math
Execution of Data/Mem math
A numeric value other than 0 (zero) is specified as the
electrical delay or phase offset.
Smoothing: ON
Conversion: ON (Impedance: Reflection measurement)
Conversion: ON (Impedance: Transmission measurement)
Conversion: ON (Admittance: Reflection measurement)
Conversion: ON (Admittance: Transmission measurement)
Conversion: ON (Inverse S-parameter)
-
. You can also move the position of the
Reference Position
to
.), moving the indicator vertically with the left mouse
Chapter 2