Specifications and Supplemental Information

Uncorrected System Performance

Uncorrected System Performance
Table 15-6
Uncorrected System Performance
Description
Directivity
Source Match
Load Match
Transmission Tracking
Reflection Tracking
334
Specification
3 MHz to 3 GHz
3 GHz to 6 GHz
10 dB
6 dB
15 dB
10 dB
17 dB
12 dB
± 3.0 dB, typical
± 2.0 dB, typical
± 3.0 dB, typical
± 2.0 dB, typical
6 GHz to 8.5 GHz
4 dB
10 dB
12 dB
± 4.0 dB, typical
± 4.0 dB, typical