Softkey
4-1 (S41)
1-2 (S12)
3-2 (S32)
4-2 (S42)
1-3 (S13)
2-3 (S23)
4-3 (S43)
1-4 (S14)
2-4 (S24)
3-4 (S34)
Step 6. Make a THRU connection between the test ports (between the connectors to which the
DUT will be connected) selected in Step 5.
Figure 4-8
Connecting at Thru Response Calibration
Step 7. Press
Chapter 4
THRU Response Calibration (Transmission Test)
Function
Selects test port 4 (input) and test port 1 (output). Corresponds to the
determination of S
Selects test port 1 (input) and test port 2 (output). Corresponds to the
determination of S
Selects test port 3 (input) and test port 2 (output). Corresponds to the
determination of S
Selects test port 4 (input) and test port 2 (output). Corresponds to the
determination of S
Selects test port 1 (input) and test port 3 (output). Corresponds to the
determination of S
Selects test port 2 (input) and test port 3 (output). Corresponds to the
determination of S
Selects test port 4 (input) and test port 3 (output). Corresponds to the
determination of S
Selects test port 1 (input) and test port 4 (output). Corresponds to the
determination of S
Selects test port 2 (input) and test port 4 (output). Corresponds to the
determination of S
Selects test port 3 (input) and test port 4 (output). Corresponds to the
determination of S
to start the calibration measurement.
Thru
.
41
.
12
.
32
.
42
.
13
.
23
.
43
.
14
.
24
.
34
Calibration