Limit Test
Limit Test
NOTE
You can define a limit line that is able to freely overlap with the frequency of another limit
line.
Defining one limit line having the same type as a second limit line whose frequency
overlaps with the first one results in two or more limit values at the same frequency point.
In this case, the limit value to be used in the limit test is defined as follows:
Display of the Limit Test Evaluation Result
During execution of the limit test, the magnitude of the measurement value and limit value
are compared with one another at all points in the limit line frequency range. Based on
whether points outside the limit value exist or not, the result of the evaluation on a
Pass/Fail basis is displayed on the screen.
Examples of the evaluation result are shown below.
Example 10-1
Acceptable Measurement Result
If all measurement values are within the limit value, as shown below, "Pass" is displayed
on the screen.
218
When two or more limit values whose type is set to maximum (MAX) exist, the
smallest one is used as the maximum.
When two or more limit values whose type is set to minimum (MIN) exist, the largest
one is used as the minimum.