Full Two-Port Error-Correction

Removes directivity errors of the test setup in forward and reverse directions.
n
Removes source match errors of the test setup in forward and reverse directions.
n
Removes load match errors of the test setup in forward and reverse directions.
n
Removes isolation errors of the test setup in forward and reverse directions (optional).
n
Removes frequency response of the test setup in forward and reverse directions.
n
This is the most accurate error-correction procedure. Since the analyzer takes
both forward and reverse sweeps, this procedure takes more time than the
other correction procedures.
1. Set any measurement parameters that you want for the device measurement: power,
format, number of points, IF bandwidth.
2. To access the measurement correction menus, press:
3. If your calibration kit is not the 3.5 mm (2.4 mm, HP 8722D) default, press:
If your type of calibration kit is not listed in the displayed menu, refer to the "Modifying
Calibration Kit Standards" procedure, located later in this chapter.
4.
5. Connect a shielded open circuit to PORT 1.
Include any adapters that you will have in the device measurement. That is,
connect the standard to the particular connector where you will connect your
device under test.
FOR REFLECTION
Figure 5-7. Standard Connections for Full Two port Error-Correction
FOR TRANSMISSION
FOR ISOLATION
Optimizing Measurement Results 5-21