Agilent Technologies ENA Series User Manual page 72

Rf network analyzer
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Setting Measurement Conditions
Stimulus Settings
Setting Up the Sweep Time (Manual Sweep Time Mode)
Step 1. Press
Step 2. Press
Step 3. Press
Sweep Time
Step 4. Using the ENTRY block keys on the front panel, input the desired sweep time (in seconds).
If the previous operation mode was automatic sweep time mode, entering a new sweep
time forces the machine to switch to manual sweep time mode.
Switching to Automatic Sweep Time Mode
Step 1. Press
automatic sweep time mode.
Step 2. Press
Step 3. Press
Sweep Time
Step 4. Press
. (By entering zero (seconds), automatic sweep time becomes effective.)
64
or
to select the channel for which sweep time will be set.
.
.
or
to select the channel which will be switched to
.
.
Chapter 3