Average Menu - Agilent Technologies ENA Series User Manual

Rf network analyzer
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Average Menu

Key Operation
Displays softkeys for setting averaging options.
Resets the counter and restarts from "1".
Averaging Restart
Sets the averaging factor. The averaging factor must be defined as an
Avg Factor
integer from 1 to 999.
Enables or disables averaging execution
Averaging
Sets the smoothing aperture using a percentage against the sweep span
Smo Aperture
Enables or disables the smoothing function. When enabled, the letters
Smoothing
"Smo" will appear in the trace status area.
Sets the IF bandwidth. For more about the effective range of the IF
IF Bandwidth
bandwidth, refer to "Specifications and Reference Data". If an input value
is out of the range, it will automatically be modified to the closest value in
the effective range. A narrow IF bandwidth slows down the sweep speed,
but improves the S/N ratio.
Returns to the "E5070A/E5071A Menu (Top Menu)" on page 404.
Return
Appendix D
Function
Softkey Functions
Average Menu
SCPI Command
:SENS{1-9}:AVER:CLE
:SENS{1-9}:AVER:COUN
:SENS{1-9}:AVER
:CALC{1-9}:SMO:APER
:CALC{1-9}:SMO:STAT
or
:SENS{1-9}:BAND
:SENS{1-9}:BWID
413