Softkey Functions

Measurement Menu (Balance Measurement, SE-Bal)

Measurement Menu (Balance Measurement, SE-Bal)
Key Operation
Displays softkeys for setting measurement parameters (only for models with Option 313, 314, 413, or 414). To
display these softkeys, the balanced/unbalanced conversion topology must be set to unbalanced-balanced
(
SE-Bal
must be on (
Selects parameter S
Sss11
signal input to (unbalanced) port 1 on the DUT is reflected as an
unbalanced signal.
Selects parameter S
Sds21
signal input to (unbalanced) port 1 on the DUT is transmitted to
(balanced) port 2 on the DUT as a differential signal.
Selects parameter S
Ssd12
signal input to (balanced) port 2 on the DUT is transmitted to
(unbalanced) port 1 on the DUT as an unbalanced signal.
Selects parameter S
Scs21
signal input to (unbalanced) port 1 on the DUT is transmitted to
(balanced) port 2 on the DUT as a common mode signal.
Selects parameter S
Ssc12
signal input to (balanced) port 2 on the DUT is transmitted to
(unbalanced) port 1 on the DUT as an unbalanced signal.
Selects parameter S
Sdd22
signal input to (balanced) port 2 on the DUT is reflected as a
differential signal.
Selects parameter S
Scd22
signal input to (balanced) port 2 on the DUT is reflected as a
common mode signal.
Selects parameter S
Sdc22
signal input to (balanced) port 2 on the DUT is reflected as a
differential signal.
Selects parameter S
Scc22
signal input to (balanced) port 2 on the DUT is reflected as a
common mode signal.
Selects parameter Imbalance.
Imbalance
Selects parameter CMRR.
CMRR
Returns to the "E5070A/E5071A Menu (Top Menu)" on page 404.
Return
436
Function
), the balanced/unbalanced conversion function must be on (
) in the "Analysis Menu" on page 405.
Fixture Simulator ON
. S
defines the way an unbalanced
ss11
ss11
. S
defines the way an unbalanced
ds21
ds21
. S
defines the way a differential
sd12
sd12
. S
defines the way an unbalanced
cs21
cs21
. S
defines the way a common mode
sc12
sc12
. S
defines the way a differential
dd22
dd22
. S
defines the way a differential
cd22
cd22
. S
defines the way a common mode
dc22
dc22
. S
defines the way a common mode
cc22
cc22
SCPI Command
), and the fixture simulator function
BalUn ON
:CALC{1-9}:FSIM:BAL:PAR{1-9}:SBAL SSS11
:CALC{1-9}:FSIM:BAL:PAR{1-9}:SBAL SDS21
:CALC{1-9}:FSIM:BAL:PAR{1-9}:SBAL SSD12
:CALC{1-9}:FSIM:BAL:PAR{1-9}:SBAL SCS21
:CALC{1-9}:FSIM:BAL:PAR{1-9}:SBAL SSC12
:CALC{1-9}:FSIM:BAL:PAR{1-9}:SBAL SDD22
:CALC{1-9}:FSIM:BAL:PAR{1-9}:SBAL SCD22
:CALC{1-9}:FSIM:BAL:PAR{1-9}:SBAL SDC22
:CALC{1-9}:FSIM:BAL:PAR{1-9}:SBAL SCC22
:CALC{1-9}:FSIM:BAL:PAR{1-9}:SBAL IMB
:CALC{1-9}:FSIM:BAL:PAR{1-9}:SBAL CMRR