Calibration
Changing the Calibration Kit Definition
Step 8. Select a type of standard.
Softkey
Open
Short
Load
Delay/Thru
Arbitrary
None
Step 9. Set the standard coefficient.
Softkey
C0
C1
C2
C3
L0
L1
L2
L3
Offset Delay
Offset Z0
Offset Loss
Arb. Impedance
Step 10. Press
and input a new label for the standard using the keypad displayed on the
Label
screen.
Step 11. Press
.
Return
Step 12. Repeat Step 6 to Step 11 to redefine all standards for which changes are necessary.
Step 13. Press
.
Return
Step 14. Press
Specify CLSs
114
Function
Selects the OPEN standard.
Selects the SHORT standard.
Selects the LOAD standard.
Selects the delay/THRU standard.
Selects the arbitrary impedance.
Selects no standard type.
Function
Sets C0.
Sets C1.
Sets C2.
Sets C3.
Sets L0.
Sets L1.
Sets L2.
Sets L3.
Sets the offset delay.
Sets the offset Z0.
Sets the offset loss.
Sets an arbitrary impedance.
.
Chapter 4