Softkey Functions
Measurement Menu (Balanced Measurement, SE-SE-Bal)
Key Operation
(Continued)
Selects parameter S
Scc33
signal input to (balanced) port 3 on the DUT is reflected as a
common mode signal.
Selects parameter Imbalance1.
Imbalance1
Selects parameter Imbalance2.
Imbalance2
Selects parameter CMRR1.
Sds31/Scs31
Selects parameter CMRR2.
Sds32/Scs32
Returns to the "E5070A/E5071A Menu (Top Menu)" on page 404.
Return
440
Function
. S
defines the way a common mode
cc33
cc33
SCPI Command
:CALC{1-9}:FSIM:BAL:PAR{1-9}:SSB SCC33
:CALC{1-9}:FSIM:BAL:PAR{1-9}:SSB IMB1
:CALC{1-9}:FSIM:BAL:PAR{1-9}:SSB IMB2
:CALC{1-9}:FSIM:BAL:PAR{1-9}:SSB CMRR
:CALC{1-9}:FSIM:BAL:PAR{1-9}:SSB CMRR2