Measurement Examples
Evaluating a Duplexer
Setup Description
Figure 14-16
Completed Limit Table for Trace 1
Step 2. Define the setup of the limit table for Trace 2 (S
Setup Description
Activating Trace 2
Segment 1
Segment 2
Segment 3
298
Begin Stimulus value: 2.13 GHz
End Stimulus value: 6.02 GHz
Begin Response value: -25 dB
End Response value: -25 dB
Type: Upper limit
Begin Stimulus value: 1.73 GHz
End Stimulus value: 1.85 GHz
Begin Response value: -40 dB
End Response value: -40 dB
Type: Upper limit
Begin Stimulus value: 1.85 GHz
End Stimulus value: 1.91 GHz
Begin Response value: -45 dB
End Response value: -45 dB
Key Operation
).
21
Key Operation
(Select
) -
Max
(Select
) -
Max
Chapter 14