Agilent Technologies ENA Series User Manual page 98

Rf network analyzer
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Calibration
THRU Response Calibration (Transmission Test)
Step 8. If an isolation calibration must be performed using a LOAD standard, follow the procedure
below.
a. Connect a LOAD standard to each of the two test ports (connectors to which the DUT
Figure 4-9
Connecting the Load Standard
b. Press
Step 9. Press
process. Upon pressing the key, calibration coefficients will be calculated and saved. The
error correction function will also be automatically enabled.
NOTE
By pressing
calibration coefficients. To cancel the calibration without saving the new calibration
coefficients, press
90
will be connected) selected in Step 5.
Isolation (Optional)
to terminate the response calibration (and the load isolation calibration)
Done
, previously saved calibration coefficients will be overwritten with new
Done
Cancel - OK
to start the calibration measurement.
successively.