Full Two-Port Error-Correction Procedure - Agilent Technologies 8753ES Option 011 Service Manual

Network analyzer
Hide thumbs Also See for 8753ES Option 011:
Table of Contents

Advertisement

Error Terms

Full Two-Port Error-Correction Procedure

Full Two-Port Error-Correction Procedure
This is the most accurate error-correction procedure for the
NOTE
analyzer. Since the analyzer takes both forward and reverse sweeps, this
procedure takes more time than the other correction procedures.
1. Set any measurement parameters that you want for the device measurement: power,
format, number of points, IF bandwidth.
2. To access the measurement correction menus, press
3. Press
CAL KIT
4. To select the correction type, press:
CALIBRATE MENU FULL 2-PORT
5. Connect a shielded open circuit to PORT 1.
Figure 11-1
Standard Connections for Full Two-Port Error-Correction
6. To measure the standard, when the displayed trace has settled, press
FORWARD: OPEN
The analyzer underlines the
7. Disconnect the open, and connect a short circuit to PORT 1.
8. To measure the device, when the displayed trace has settled, press
FORWARD: SHORT
The analyzer underlines the
11-4
SELECT CAL KIT
7 mm 85031
.
softkey after it measures the standard.
OPEN
.
SHORT
.
Cal
RETURN
REFLECTION
softkey after it measures the standard.
.
Chapter 11