Key Operation
(Continued)
Displays softkeys for maintenance services.
Service Menu
Displays softkeys for self diagnosis functions.
Test Menu
Executes an internal test.
Power On Test
Executes a display test.
Display
Tests the front panel keys (hardkeys).
Front Panel
Executes calibration of the touch screen.
Adjust Touch
Screen
Returns to the softkey display screen one level higher.
Return
Turns ON/OFF error correction, which uses system calibration data. If
System Correction
user calibration using [Cal] is executed properly and the error
correction is valid, you can turn off system error correction and reduce
measurement time.
Turns ON/OFF the avoid spurious mode.
Avoid Spurious
Turns ON/OFF the high temperature measurement mode. When the
High Temperature
high temperature measurement mode is turned on, drift error can be
reduced in the ambient temperature range of 28°C to 33°C. If you use
the analyzer within the range of 23°C ± 5°C, you must turn off this
function.
Displays softkeys for other options.
Enable Options
Displays softkeys for rebooting the analyzer.
Restart Menu
Displays softkeys for performing service functions. This option is not open to average users.
Service Functions
Returns to the softkey display screen one level higher.
Return
Returns to the "E5070A/E5071A Menu (Top Menu)" on page 404.
Return
*1.Only with Options 016.
Appendix D
Function
Softkey Functions
System Menu
SCPI Command
None
None
None
None
:SYST:CORR
:SENS{1-9}:SWE:ASP
:SYST:TEMP:HIGH
451