Calibration
Full 3-Port and Full 4-Port Calibration Using the 2-Port ECal
4-Port
Step 8. When a full 3-port calibration is carried out on an E5070A/E5071A with option 413 or
414, select the test ports to be calibrated on the drop down list box below the
(either
Step 9. In the Select Channel area, select the channel to be calibrated (one of the channels
Step 10. Press the
A dialog box as shown in Figure 4-22 appears.
Figure 4-22
ECalAssistant (Connection) Dialog Box
Step 11. Following the connection diagram shown in the dialog box (Figure 4-22), connect port A
and B of the ECal module to test ports on the E5070A/E5071A.
Note that the connection diagram shown in each dialog box that appears in each step
depends on the number of test ports on the E5070A/E5071A (option 313/314 or 413/414),
and the test port selection made in Step 8.
Step 12. Press the
Upon completion of the measurement, a dialog box as shown in Figure 4-23 will be
displayed.
Figure 4-23
ECalAssistant (Measurement Complete) Dialog Box
Step 13. Press the
A dialog box as shown in Figure 4-24 appears.
106
(for a full 4-port calibration) radio button.
,
,
, or
1-2-3
1-2-4
1-3-4
2-3-4
button.
Next
button to start the measurement of calibration data.
Measure
button.
Next
).
button
3-Port
~
).
1
9
Chapter 4