Measurement Examples
Measuring parameters with cable

5. Setting the time domain function

Set the conversion function to display the response in time domain. If you enable this
setting, the response in time domain is displayed as shown in Figure 14-34. A peak
indicating a small mismatch appears at the location of the connector.
Description of setting
Data format: real
Setting the transformation type to low pass
impulse
Setting the window type to maximum.
Setting the display range to from 0 s to 10 ns
Enabling the transformation function
Executing auto scale
Figure 14-34
Response in time domain
318
Key stroke
- Real
- Transform - Lowpass Imp.
Window - Maximum
Start -
Stop -
Transform (set to ON)
- Auto Scale