Analysis in Time Domain (Option 010)
Transformation to time domain
Impulse signal and step signal
The E5070A/E5071A lets you simulate the response from the DUT to 2 types of signals:
impulse signal and step signal. The impulse signal is a pulse-shape signal in which the
voltage rises from 0 to a certain value and returns to 0 again. The pulse width depends on
the frequency sweep range. The step signal is a signal in which the voltage rises from 0 to a
certain value. The rise time depends on the maximum frequency within the frequency
sweep range.
NOTE
For more information on how the frequency span setting affects the pulse width and the
rise time, refer to "Calculating necessary measurement conditions" on page 182.
Figure 8-2
Step signal and impulse signal
Operation
Step 1. Press
which you want to set the transformation type.
Step 2.
Press
Step 3.
Type
Softkey
Bandpass
Lowpass Step
Lowpass Imp.
Step 4. Press
180
(or
- Press
to display the "Transform" menu.
Transform
and then press one of the following softkeys to specify the type.
Function
Sets the transformation type to "band pass."
Sets the transformation type to "lowpass step."
Sets the transformation type to "lowpass impulse."
to display the "Format" menu and then select the data format.
) and
(or
) to activate a trace for
8