Full 2-Port Calibration Using the 2-Port ECal Module

Follow the procedure below to perform a full 2-port calibration using the 2-
port ECal module.
1. Connect the USB port on the ECal module with the USB port on the
E5071C via a USB cable. This connection may be done while the
E5071C's power is on.
2. Allows the ECal module to warm up for 15 minutes until the module
indicates from WAIT to READY.
3. Connect port A and port B on the ECal module to the test ports to be
calibrated.
4. Press
Channel Next/Channel Prev
you want to perform the calibration.
5. Press
Cal
6. Click ECal.
7. Click 2 Port ECal. When using a 2-port E5071C, pressing this key
performs a 2-port ECal. When using a 3-port or 4-port E5071C, click
one of the softkeys to start a full 2-port calibration.
Connecting ECal module (full 2-port calibration)
key.
keys to select the channel for which
Measurement