Calibration
Measurement Errors and their Characteristics
Transmission tracking error (Et)
A transmission tracking error is caused because the difference in frequency response
between the test receiver of a response port and the reference receiver of a stimulus port in
transmission measurement. When a certain port is a stimulus port, a transmission tracking
error is defined for each of the other ports. Therefore, the number of transmission tracking
errors of the E5070A/E5071A is the total number of the combinations of stimulus ports and
response ports.
*1
*2
Et21, Et31
, and Et41
*1
*2
Et12, Et32
, and Et42
*1
*1
*2
Et13
, Et23
, and Et43
*2
*2
*2
Et14
, Et24
, and Et34
*1.Options 313, 314, 413, and 414 only
*2.Options 413 and 414 only
82
Transmission tracking error when port 1 is a stimulus port.
Transmission tracking error when port 2 is a stimulus port.
Transmission tracking error when port 3 is a stimulus port.
Transmission tracking error when port 4 is a stimulus port.