Measurement Examples
Evaluating transmission characteristics of a front end module

6. Setting balance conversion topology

For channel 2, set the balance conversion topology in order to perform measurement
including the balanced port.
Description of setting
Set channel 2 to the active channel.
Set DUT port 1 to unbalance and DUT port 2 to
balance.
Set the destination to which DUT port 1
(unbalance) is connected to test port 1 of the
analyzer.
Set the destination to which DUT port 2
(balance) is connected to test ports 3 and 4 of the
analyzer.
Unbalance-balance conversion for trace 1: ON
Fixture simulator: ON

7. Selecting measurement parameter

Set the measurement parameter for channel 1.
Step 1. Select the measurement parameter for trace 1 of channel 1.
Description of setting
Set channel 1 to the active channel.
Measurement parameter for trace 1: S12
NOTE
The subscript of the measurement parameter means the test port of the E5070A/E5071A.
Check the test port assignment and select the measurement parameter.
Step 2. Set the measurement parameter for channels 2, 3, and 4 according to the same procedure as
in Step 1.

8. Executing calibration

Step 1. Display the Ecal menu.
Description of setting
Display the ECal menu.
322
Key stroke
(Press it until channel 2 is
activated.)
- Fixture Simulator - Topology -
Device - SE-Bal
-
Port 1 (se)
1
-
Port 2 (bal)
3-4
(Set it to
.)
BalUn
ON
(Set it to
Fixture Simulator
ON
Key stroke
(Press it until channel 1 is
activated.)
-
S12
Key stroke
- ECal
.)