Measurement Examples
Measuring the SAW Bandpass Filter Using the Segment Sweep

2. Create a Segment Sweep Table

Follow the steps below to make entries in the segment sweep table.
Step 1. Display the segment table.
Setup Description
Presetting
Displaying the segment table
Step 2. Display the IF bandwidth setting column on the segment table.
Setup Description
Moving the focus to the softkey menu
Display of the IF bandwidth setting column:
ON
NOTE
When setup items (power level, delay time, and sweep time in this case) are not displayed
in the segment table, the setting for the channel in use applies to all segments.
Step 3. Enter the setup data into the segment table.
Setup Description
Moving the focus to the segment table
Segment 1
Segment 2
Segment 3
282
Start frequency: 440 MHz
Stop frequency: 915 MHz
Number of points: 47
IF bandwidth: 70 kHz
Start frequency: 915 MHz
Stop frequency: 980 MHz
Number of points: 130
IF bandwidth: 100 kHz
Start frequency: 980 MHz
Stop frequency: 1.035 GHz
Key Operation
-