Agilent Technologies ENA Series User Manual page 425

Rf network analyzer
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Key Operation
(Continued)
Calibrate
2-Port Cal
Terminates a calibration process and calculates the calibration
Done
coefficients.
Displays softkeys for canceling a calibration.
Cancel
Cancels the calibration in progress.
OK
Returns to the softkey display screen one level higher.
Cancel
Returns to the softkey display screen one level higher.
Return
Displays softkeys for executing full 3-port calibrations.
*1
3-Port Cal
*2
Displays softkeys for selecting test ports.
Select Ports
Selects test ports 1, 2, and 3.
1-2-3
Selects test ports 1, 2, and 4.
1-2-4
Selects test ports 1, 3, and 4.
1-3-4
Selects test ports 2, 3, and 4.
2-3-4
Returns to the softkey display screen one level higher.
Cancel
Displays softkeys for executing reflection calibrations. Symbols x, y, and z refer to the three test ports selected in
Reflection
the
Select Ports
Executes an OPEN reflection calibration on test port x.
Port x Open
Executes a SHORT reflection calibration on test port x.
Port x Short
Executes a LOAD reflection calibration on test port x.
Port x Load
Executes an OPEN reflection calibration on test port y.
Port y Open
Executes a SHORT reflection calibration on test port y.
Port y Short
Executes a LOAD reflection calibration on test port y.
Port y Load
Executes an OPEN reflection calibration on test port z.
Port z Open
Executes a SHORT reflection calibration on test port z.
Port z Short
Executes a LOAD reflection calibration on test port z.
Port z Load
Returns to the softkey display screen one level higher.
Return
Displays softkeys for executing transmission calibrations. Symbols x, y, and z refer to the three test ports selected
Transmission
in the
Select Ports
Executes a THRU calibration on test ports x and y.
Port x-y Thru
Executes a THRU calibration on test ports x and z.
Port x-z Thru
Executes a THRU calibration on test ports y and z.
Port y-z Thru
Returns to the softkey display screen one level higher.
Return
Displays softkeys for executing isolation calibrations. Symbols x, y, and z refer to the three test ports selected in
Isolation
the
(Optional)
Select Ports
Executes an isolation test on test ports x and y.
Port x-y Isol
Executes an isolation test on test ports x and z.
Port x-z Isol
Executes an isolation test on test ports y and z.
Port y-z Isol
Returns to the softkey display screen one level higher.
Return
Appendix D
Function
menu (x, y, and z are always 1, 2, and 3, respectively, for models with Option 313 or 314).
menu (x, y, and z are always 1, 2, and 3, respectively, for models with Option 313 or 314).
menu (x, y, and z are always 1, 2, and 3, respectively, for models with Option 313 or 314).
Softkey Functions
Calibration Menu
SCPI Command
:SENS{1-9}:CORR:COLL:SAVE
None
:SENS{1-9}:CORR:COLL:METH:SOLT3 1,2,3
:SENS{1-9}:CORR:COLL:METH:SOLT3 1,2 4
:SENS{1-9}:CORR:COLL:METH:SOLT3 1,3,4
:SENS{1-9}:CORR:COLL:METH:SOLT3 2,3,4
:SENS{1-9}:CORR:COLL:OPEN x
:SENS{1-9}:CORR:COLL:SHOR x
:SENS{1-9}:CORR:COLL:LOAD x
:SENS{1-9}:CORR:COLL:OPEN y
:SENS{1-9}:CORR:COLL:SHOR y
:SENS{1-9}:CORR:COLL:LOAD y
:SENS{1-9}:CORR:COLL:OPEN z
:SENS{1-9}:CORR:COLL:SHOR z
:SENS{1-9}:CORR:COLL:LOAD z
:SENS{1-9}:CORR:COLL:THRU x,y
:SENS{1-9}:CORR:COLL:THRU y,x
:SENS{1-9}:CORR:COLL:THRU x,z
:SENS{1-9}:CORR:COLL:THRU z,x
:SENS{1-9}:CORR:COLL:THRU y,z
:SENS{1-9}:CORR:COLL:THRU z,y
:SENS{1-9}:CORR:COLL:ISOL x,y
:SENS{1-9}:CORR:COLL:ISOL y,x
:SENS{1-9}:CORR:COLL:ISOL x,z
:SENS{1-9}:CORR:COLL:ISOL z,x
:SENS{1-9}:CORR:COLL:ISOL y,z
:SENS{1-9}:CORR:COLL:ISOL z,y
417