Scale Menu - Agilent Technologies ENA Series User Manual

Rf network analyzer
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Softkey Functions

Scale Menu

Scale Menu
Key Operation
Displays softkeys for adjusting scales.
Automatically adjusts scales for the active trace.
Auto Scale
Automatically adjusts scales for all traces within the active channel.
Auto Scale All
Defines the number of divisions on the Y-axis of a rectangular display
Divisions
format. An even number from 4 to 30 must be used. Once set, it is then
applied to all traces displayed in any rectangular display format within that
channel.
In a rectangular format, defines the number of increments per division on
Scale/Div
the Y-axis. In the Smith chart format or polar format, defines the range (the
displacement of the outermost circle). The setting applies to the active trace
only.
Defines the position of the reference line on a rectangular display format.
Reference Position
The position must be defined by using numbers assigned to the Y-axis
between 0 (the origin, the X-axis) and the number of divisions (the highest
scale). This setting applies only to the active trace. The reference line can
also be moved by performing a drag-and-drop operation (pressing the
mouse button on the object to be moved and releasing the button after
dragging it to the desired position) on one of the reference line pointers (
and
).
Defines the value corresponding to the reference line on a rectangular
Reference Value
display format. The setting applies only to the active trace.
Changes the value of the reference line to the response value of the active
Marker ® Reference
marker. The same function is also accessible from the "Marker Function
Menu" on page 432.
Sets an electrical delay to the active trace. This function simulates the
Electrical Delay
addition or deletion of a variable length lossless transfer line against the
input of a receiver. It can be used to compensate for the electrical length of
cables inside the DUT. Although the unit used is seconds, the length
(meters) will be displayed in parentheses next to the input box based on the
velocity coefficient used at the time.
Sets the values to be added or subtracted in phase measurement (phase
Phase Offset
offset) (°).
Returns to the "E5070A/E5071A Menu (Top Menu)" on page 404.
Return
444
Function
SCPI Command
:DISP:WIND{1-9}:TRAC{1-9}:Y:AUTO
None
:DISP:WIND{1-9}:TRAC{1-9}:Y:DIV
:DISP:WIND{1-9}:TRAC{1-9}:Y:PDIV
:DISP:WIND{1-9}:TRAC{1-9}:Y:RPOS
:DISP:WIND{1-9}:TRAC{1-9}:Y:RLEV
:CALC{1-9}:MARK{1-10}:SET
:CALC{1-9}:CORR:EDEL:TIME
:CALC{1-9}:CORR:OFFS:PHAS
Appendix D

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E5070E5071a

Table of Contents