Specifications and Supplemental Information
Corrected System Performance
Table 15-2
Corrected System Performance With Type-N Device Connectors, 85032F
Calibration Kit
Network analyzer: E5070A/E5071A, Calibration kit: 85032F (Type-N, 50 W), Calibration: full 2-port
IF bandwidth = 10 Hz, No averaging applied to data, Environmental temperature = 23°C±5°C with <1°C deviation
from calibration temperature, Isolation calibration not omitted
Description
Directivity
Source Match
Load Match
Reflection Tracking
Transmission Tracking
Transmission Uncertainty (Specification)
Reflection Uncertainty (Specification)
330
3 MHz to 3 GHz
49
41
49
±0.011
±0.016
Specification (dB)
3 GHz to 6 GHz
40
36
40
±0.032
±0.062
6 GHz to 8.5 GHz
38
35
38
±0.054
±0.073
Chapter 15