5 Panoramic and 3D calibrations
The mechanism upper mechanical limit must be approx. 10 mm (0.4 in.)
below the upmost position (in the calibration mode).
NOTE
If the X-ray unit is equipped with ProFace sensor, the limit must be
approx. 20 mm (0.8 in.).
Technical manual
Planmeca ProMax 3D Max