8 Parts replacement and repair

8.1.3 Removing C-arm inner cover

8.1.4 Removing tube head covers (96 kV tube)

194
Planmeca ProMax 3D Max
NOTE
You have to remove the sensor covers first.
1. Unscrew the four screws that attach the inner cover to the C-arm (1), see
the figure below.
2. Remove the inner cover (2).
NOTE
If the X-ray unit is equipped with optional 120 kV X-ray tube, see instructions
given in section "Removing tube head covers (120 kV tube)" on page 196.