Network Settings Reset; Resetting X-Ray Unit Ip Address - Planmeca ProMax 3D Max with ProTouch Technical Manual

Hide thumbs Also See for ProMax 3D Max with ProTouch:
Table of Contents

Advertisement

2.3.4 Network settings reset

2.3.5 Resetting X-ray unit IP address

Technical manual
Select Network Settings Reset.
Touch the Reset button to reset the display unit (ProTouch GUI) IP address
to its default value of 10.0.0.10.
NOTE
There should not be any reason to change the GUI IP address in the first
place. If, however, the GUI IP address has been modified, in some rare
situations resetting the GUI IP address might be required. If the GUI display
device is replaced, for example, there might be a communication failure
between the X-ray unit and the new GUI device.
If connection to the X-ray unit is lost because of unknown IP address, and
for some reason you are not able to check the IP address from the Network
Settings (1400) information menu, it is possible to reset the IP address.
To reset the IP address of the X-ray unit, follow the steps below.
1. Switch off the X-ray unit.
2 User interface
Planmeca ProMax 3D Max
45

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ProMax 3D Max with ProTouch and is the answer not in the manual?

Questions and answers

Table of Contents

Save PDF