1 General and technical data
1.5.11.4 The nominal X-ray voltage together with the highest X-ray tube current
obtainable from the high-voltage generator when operated at its nominal X-ray tube
voltage
1.5.11.5 The highest X-ray tube current together with the highest X-ray tube voltage
obtainable from the high-voltage generator when operated at its highest X-ray tube
current
1.5.11.6 The X-ray tube voltage and X-ray tube current which result in the highest
electric output power
1.5.11.7 The nominal electric power for a load time of 0.1 s and at the nominal X-ray
tube voltage
1.5.11.8 The nominal electric power for a load time of 4 s and at the nominal X-ray tube
voltage
1.5.11.9 The reference current time product
1.5.11.10 Nominal anode input power of the X-ray tube
1.5.11.11 Maximum anode heat content of the X-ray tube
10
Planmeca ProMax 3D Max
D-067SB: 92 kV 12 mA (3D mode)
D-059SBR: 120 kV 8 mA (3D mode)
SXR 130-10-0.5 SC: 120 kV 11 mA (3D mode)
12 mA 96 kV
96 kV 12 mA
96 kV 12 mA - 1152 W
D-067SB: 96 kV 12 mA - 1152 W
D-059SBR: 120 kV 10 mA - 1200 W
SXR 130-10-0.5 SC: 120 kV 11 mA - 1320 W
1.2 mAs
Irradiation time 0.2 s, X-ray tube current 6 mA, X-ray tube voltage: whole
range
1152 W
D-067SB: 35 kJ
D-059SBR: 28 kJ
SXR 130-10-0.5 SC: 31.5 kJ