Prostyle Intra
USER'S MANUAL
En
Publication part number 688092
Version 20
Published 2001-08

Advertisement

Table of Contents
loading

Summary of Contents for Planmeca Prostyle Intra

  • Page 1 Prostyle Intra USER'S MANUAL Publication part number 688092 Version 20 Published 2001-08...
  • Page 3: Table Of Contents

    Programming the preprogrammed settings ....................20 MAINTENANCE ........................... 22 11.1 Cleaning ..............................22 11.2 Preventive maintenance ..........................22 EXPOSURE VALUE TABLES ........................23 ERROR CODES ............................ 25 DISPOSAL OF THE UNIT ........................26 TECHNICAL SPECIFICATIONS ......................27 User's manual Prostyle Intra TOC - 1...
  • Page 4 IEC364 - equipment is used according to the operating instructions. Planmeca pursues a policy of continual product development. Although every effort is made to produce up-to-date product documentation this publication should not be regarded as an infallible guide to current specifications. We re- serve the right to make changes without prior notice.
  • Page 5: Introduction

    NOTE This manual is valid for software revisions 1.20 or la- ter. NOTE The use of Prostyle Intra x-ray unit is allowed only under supervision of dental care professional. CAUTION The x-ray unit may be dangerous to both patient and operator unless safe exposure values are used and correct operating procedures are observed.
  • Page 6: Symbols

    SYMBOLS SYMBOLS Type B equipment (Standard IEC 601-1). Alternating current (Standard IEC-417). Attention, consult accompanying documents (Standard IEC 601-1). Intermediate focal spot (Standard IEC-417). Radiation filter or filtration (Standard IEC-417). User's manual Prostyle Intra X-ray Unit...
  • Page 7: Check List Before Using The X-Ray Unit

    • Make sure that you are using the correct film process- ing chemicals for the film you are using. • Make sure that the processing chemicals you are using are fresh and are at the correct processing temperatures and concentrations. Prostyle Intra X-ray Unit 3 User's manual...
  • Page 8: Main Parts

    One end of the control panel cable is connected to the terminal at the under-side of the generator box, and the other end to the control panel. CAUTION Do not connect any other equipment to the control panel’s terminal. User's manual Prostyle Intra X-ray Unit...
  • Page 9 0,5 - 1 rounds counterclockwise to loosen it. NOTE Do not turn the adjustment screw too much counter- clockwise to avoid the screw to come loose. Horizontal tube head stiffness adjustment screw Prostyle Intra X-ray Unit 5 User's manual...
  • Page 10: Control Panel

    In the digital imaging mode the exposure time is c . 2 0 shown with prefix c, and the exposure time is always less than one second, i.e. the maximum exposure time is 0.80 seconds. User's manual Prostyle Intra X-ray Unit...
  • Page 11: Keys And Indicator Lights

    All the keys function as in the nor- mal mode, only the density scale (0 ... -9) is different c . 2 0 than in the normal mode (5 ... -5). Prostyle Intra X-ray Unit 7 User's manual...
  • Page 12 Exposure warning indicator light The exposure warning light will come on when you ta- ke an exposure. You will also hear an audible warning sound during the exposure. User's manual Prostyle Intra X-ray Unit...
  • Page 13: Preparations For The Exposure

    Attach/remove: red points in line. In position: red and black points in line. Prostyle Intra X-ray Unit 9 User's manual...
  • Page 14 NOTE The RINN film holder can be used only with the green RINN compatible rectangular collimator. Attach the RINN compatible rectangular collimator to the long cone and then attach the film holder to the col- limator. RINN compatible rectangular collimator RINN film holder User's manual Prostyle Intra X-ray Unit...
  • Page 15 (2). The cone can be now ro- tated in its position ±90°. 180° Red points The rectangular cone can be removed when the red dots on the short cone and on the rectangular cone are in line. Prostyle Intra X-ray Unit 11 User's manual...
  • Page 16: Selecting The Exposure Parameters

    NOTE After adjusting the kV value the unit will return auto- matically to the time adjustment mode after 5 seconds time. User's manual Prostyle Intra X-ray Unit...
  • Page 17: Positioning The Patient

    The film or sensor is placed to a film holder which is used to align the film parallel to the long axis of the tooth. Film or Long axis sensor of the tooth Prostyle Intra X-ray Unit 13 User's manual...
  • Page 18: Positioning The Cone

    Table 1:Angle of inclination of the exposure cone to the horizontal plane Angle of Teeth inclination Maxilla Molars +35° Premolars and +45° canine teeth Incisors +55° Occlusal exposure +75° Mandible Occlusal exposure -60° Incisors -20° Premolars and -10° canine teeth Molars -5° Bite-wing exposure 5° User's manual Prostyle Intra X-ray Unit...
  • Page 19 POSITIONING THE PATIENT Position the cone according to the figures below. Maxillary anterior Maxillary molar Mandibular anterior Mandibular canine Film Mandibular molar Mandibular occlusal Film Maxillary occlusal Bite-wing Prostyle Intra X-ray Unit 15 User's manual...
  • Page 20: Taking An Exposure

    Press and hold the exposure key on the control panel ATTENTION: RAYONS-X for the duration of the exposure. READY PRET 8 mA SELECT The exposure warning light will come on. You will al- so hear an audible warning sound during the exposure. User's manual Prostyle Intra X-ray Unit...
  • Page 21: Exposure Values

    Table 3:Default exposure values with 30 cm (12”) cone PATIENT time Adult 0.40 Child 0.32 NOTE When using a film holder select 1-2 steps longer ex- posure time. Prostyle Intra X-ray Unit 17 User's manual...
  • Page 22: Preprogrammed Settings Values

    0.12 0.16 0.20 0.16 Mandible 0.10 0.12 0.16 0.16 When using the 30 cm long cone program the values according to the table given in section 12 “EXPO- SURE VALUE TABLES” on page 23. User's manual Prostyle Intra X-ray Unit...
  • Page 23: Programming The Exposure Values

    When the select key is pressed a second time briefly, the SELECT current density value starts to flash on the time display and the kV value is stored in the memory. c - 4 Prostyle Intra X-ray Unit 19 User's manual...
  • Page 24: Programming The Preprogrammed Settings

    The indicator light of the se- lected projection will come on. 8 mA The current time and kV values appear on the respective displays. c . 2 0 0 . 2 0 User's manual Prostyle Intra X-ray Unit...
  • Page 25 The new kV value is stored in the memory. NOTE If you interrupt programming for over 45 seconds, the unit automatically exits the programming mode, and the current values will be stored in the memory. Prostyle Intra X-ray Unit 21 User's manual...
  • Page 26: Maintenance

    11.2 Preventive maintenance To guarantee operator and patient safety and to ensure image quality the unit must be checked by a qualified service technician once a year or after every 10,000 exposures if this is sooner. User's manual Prostyle Intra X-ray Unit...
  • Page 27: Exposure Value Tables

    MAXILLA adult MANDIBLE 63 kV/ MAXILLA adult MANDIBLE MAXILLA 60 kV/ adult MANDIBLE 57 kV/ MAXILLA adult MANDIBLE MAXILLA 55 kV/ adult MANDIBLE 52 kV/ MAXILLA adult MANDIBLE 50 kV/ MAXILLA adult MANDIBLE User's manual Prostyle Intra X-ray Unit 23...
  • Page 28 55 kV/ MAXILLA adult MANDIBLE 52 kV/ MAXILLA adult MANDIBLE 50 kV/ MAXILLA adult MANDIBLE INCISORS PREMOLARS AND CANINES MOLARS OCCLUSAL EXPOSURE NOTE When using the film holder select 1-2 steps longer ex- posure times. User's manual Prostyle Intra X-ray Unit...
  • Page 29: Error Codes

    Communication error between Display and Tube head CPU. E . 7 1 FLASH memory check-sum error (Tube head CPU). E . 8 1 EEPROM memory defective (Tube head CPU). E . 8 3 Config register error (Tube head CPU). Prostyle Intra X-ray Unit 25 User's manual...
  • Page 30: Disposal Of The Unit

    DISPOSAL OF THE UNIT DISPOSAL OF THE UNIT In order to reduce the environmental load over the pro- duct’s entire lifecycle, PLANMECA’s products are desig- ned to be as safe as possible to manufacture, use and dispo- se of. Parts which can be recycled should always be taken to the appropriate processing centres, after hazardous waste has been removed.
  • Page 31: Technical Specifications

    1:15, automatic control Electrical classification Class I, type B Weight total 23 kg (51 lbs) tube head 4.5 kg (10 lbs) Ambient temperature operating +5°C - +40°C storage -10°C - +50°C Humidity 25% - 75% Prostyle Intra X-ray Unit 27 User's manual...
  • Page 32 TECHNICAL SPECIFICATIONS Dimensions (in mm) 1790 (70.5") (3") 535 (std) (21.1") 335 (opt) (13.2") 835 (opt) (32.9") (4.25") (6.25") 1470-1670-1970 (57.5"-65.5"-77") User's manual Prostyle Intra X-ray Unit...
  • Page 33 Asentajankatu 6 Egedal 1 C 00810 Helsinki 2690 Karlslunde Finland Tel. +45 46 155 251 Tel. +358 9 759 05500 Planmeca Middle East Fax + 358-9-759 05 555 www.planmeca.com Al-Moosa Tower II, e-mail: sales@planmeca.com Sheikh Zayed Road P.O. Box 28826...