2.5.9 Changing X-ray unit mode

Steps
Technical manual
3. Modify the settings in the Yearly Maintenance Reminder display.
To activate the reminder, set the Yearly Maintenance Reminder
button on the left to ON and to deactivate set the button to OFF.
By default the setting is deactivated.
With the up/down button, set the reminder interval in months.
The maximum allowed value is 24 months, minimum 6 months and
default is 11 months.
The maintenance reminder appearance date is displayed in the
lower left corner.
The Reset button resets the maintenance interval to the default 11
months.
4. Touch the green check mark button to activate the changes and leave
the settings display.
When the maintenance reminder is activated, the help message H110 is
displayed after every boot up a after the service due date.
1. Enter the technical settings menu.
See more in section "Entering technical settings menu" on page 51.
2 User interface
Planmeca ProMax 3D Max
61