3 Help and error messages
Error
Explanation
code
E401
Tube head offsets
E402
E403
E404
E405
E406
E407
E408
84
Planmeca ProMax 3D Max
Tube head KVPOS offset failure, short-
circuited.
Tube head KVPOS offset failure, out of
bounds.
Tube head offset KVNEG failure, short-
circuited.
Tube head KVNEG offset failure, out of
bounds.
Tube head MAPOS offset failure, short-
circuited.
Tube head MAPOS offset failure, out of
bounds.
Tube head MANEG offset failure, short-
circuited.
Tube head MANEG offset failure, out of
bounds.
Comments
Exposure not
possible. Check the
cabling between the
FBK and the CAM
PCB. Check also the
functionality of the
FBK PCB.
Exposure not
possible. Check the
cabling between the
FBK and the CAM
PCB. Check also the
functionality of the
FBK PCB.
Exposure not
possible. Check the
cabling between the
FBK and the CAM
PCB. Check also the
functionality of the
FBK PCB.
Exposure not
possible. Check the
cabling between the
FBK and the CAM
PCB. Check also the
functionality of the
FBK PCB.
Exposure not
possible. Check the
cabling between the
FBK and the CAM
PCB. Check also the
functionality of the
FBK PCB.
Exposure not
possible. Check the
cabling between the
FBK and the CAM
PCB. Check also the
functionality of the
FBK PCB.
Exposure not
possible. Check the
functionality of the
FBK PCB.
Exposure not
possible. Check the
functionality of the
FBK PCB.
Technical manual
Need help?
Do you have a question about the ProMax 3D Max with ProTouch and is the answer not in the manual?
Questions and answers