Planmeca 2D S2 User Manual

Planmeca 2D S2 User Manual

Dental equipment/2d/3d imaging
Hide thumbs Also See for 2D S2:
Table of Contents

Advertisement

Planmeca ProMax
®
2D & 3D s & 3D Classic
user's manual
2D imaging

Advertisement

Table of Contents
loading

Summary of Contents for Planmeca 2D S2

  • Page 1 Planmeca ProMax ® 2D & 3D s & 3D Classic user's manual 2D imaging...
  • Page 3: Table Of Contents

    Selecting exposure settings ....................33 Patient positioning ........................ 36 Selecting jaw size and shape ....................40 Selecting MultiView (Planmeca ProMax 3D X-ray units) ............41 Adjusting exposure values for current exposure ..............42 Selecting Dynamic Exposure Control (DEC) (Planmeca ProMax 2D X-ray units) ..................44 Taking an exposure ......................
  • Page 4 IEC 60364 - equipment is used according to the operating instructions. Planmeca pursues a policy of continual product development. Although every effort is made to produce up-to-date product documentation this publication should not be regarded as an infallible guide to current specifications. We reserve the right to make changes without prior notice.
  • Page 5: Introduction

    The X-ray unit uses panoramic techniques to produce two-dimensional (2D) X-ray images for the examination of dentomaxillofacial anatomy. You need a PC with the Planmeca Romexis program in order to save, view and modify the images. Make sure that you are fully acquainted with the appropriate radiation protection measures and these instructions before you use the X-ray unit.
  • Page 6: Associated Documentation

    The User’s Manuals are available on Planmeca’s website. • For X-ray units, select Material bank > Manuals > Imaging. • For software products, select Material bank > Manuals > Software. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 7: Symbols On Product Labels

    Emergency stop (Standard IEC 60601-1) Warning; Electricity (Standard ISO 7010-W012) Electrostatic sensitive device (Standard IEC 60417) Warning; Hot surface (Standard ISO 7010-W017) General warning sign (Standard ISO 7010-W001) User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 8: Symbols On X-Ray Units With Detachable Power Supply Cord

    (Directive 2002/96/EC WEEE) Attention, consult accompanying documents (Standard IEC 60601-1) Alternating current (Standard IEC 60417) Electrostatic sensitive device (Standard IEC 60417) Warning; Hot surface (Standard ISO 7010-W017) Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 9: Safety Precautions

    CAUTION If the X-ray unit shows any signs of oil leakage, switch the X-ray unit off and contact your service technician for help. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 10 If the X-ray system is not connected to an Uninterruptible Power Supply (UPS), switch the X-ray unit off and disconnect the PCs from the mains during lightning storms. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 11 Contact your service technician if you notice a decrease in image quality. NOTE Contact your service technician if you have taken an exposure but the image does not appear in the Planmeca Romexis program. The last ten images can be manually imported into Romexis. NOTE Do not handle liquids near or on the X-ray unit.
  • Page 12 Patients are not allowed to hang on the patient handles. NOTE FOR PROFACE SENSOR: Do not touch the glass windows. Fingerprints or other stains on the glass surface destroy image quality. Prof_6C.eps Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 13: Switching X-Ray Unit On

    The on / off switch is located on the underside of the stationary column top. NOTE To prolong the lifetime of your X-ray unit, always switch the X-ray unit off when it is not in active use. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 14: Main Parts

    6.1 General view of 2D X-ray system 2 & 3 Ethernet X-ray unit Planmeca Romexis program ProTouch desktop application (optional, see section 6.10 “ProTouch desktop application” on page 19) Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 15: General View Of 3D X-Ray System

    6.2 General view of 3D X-ray system 3 & 4 Ethernet X-ray unit 3D reconstruction PC Planmeca Romexis program ProTouch desktop application (optional, see section 6.10 “ProTouch desktop application” on page 19) User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 16: General View Of X-Ray Unit

    (see section 6.9 “Touch screen” on page 16) Telescopic column (see section 6.6 “Telescopic column” on page 14) Stationary column 10 Emergency stop button (see section 6.8 “Emergency stop button” on page 15) Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 17: Sensor

    Dimax sensor 3D sensor for Planmeca ProMax 3D s 3D sensor for Planmeca ProMax 3D Classic ProFace sensor for Planmeca ProMax 3D s and Planmeca ProMax 3D Classic User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 18: Patient Supports

    Bite piece Chin rest Adapter Chin cup Chin support Temple supports 6.6 Telescopic column Telescopic column without accessory cabinet Telescopic column with accessory cabinet (three accessory boxes) Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 19: Exposure Switch

    A help message will appear on the touch screen. Guide the patient away from the X-ray unit. Then release the emergency stop button. The X-ray unit will automatically restart. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 20: Touch Screen

    • Program group button(s): To view all the programs that are available for a program group, select the program group button. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 21 • Moving backward: To go back to a previous screen, select a previous symbol at the bottom of the screen. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 22 You can switch demo mode on if you wish to practice or demonstrate the functions of the X-ray unit without radiation (Settings > User > 1300 Operational settings > 1310 Use Mode > 1311 Set Demo Mode). Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 23: Protouch Desktop Application

    To move the control panel screen, drag it with your mouse. To close the application, click on the small red cross in the top right corner of the screen. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 24: Patient Positioning Controls

    If something is in danger of becoming trapped, release the button immediately to stop the movement. Clear any obstruction before pressing the button again. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 25 Open / close temple supports Press the temple support button to open the temple supports in 2D imaging. The temple supports can be closed by pressing the temple support button again. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 26: Programs

    NOTE This program is optimized for interproximal imaging and a shadow of the opposite side teeth may therefore be visible in the image. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 27: Temporomandibular Joint (Tmj) Programs

    Planmeca ProMax 2D S3 & Planmeca ProMax 3D: The imaging position and angle can be adjusted. The default imaging angle is 0°. • Planmeca ProMax 2D S2: The imaging position can be adjusted. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 28 Three lateral multi-angle TMJ exposures (left or right). The imaging angle for image no. 2 is adjustable (three imaging angles: 0° ± 7° by default). The selected imaging angle is in image no. 2. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 29: Sinus Programs

    0° ± 7° by default). The selected imaging angle is in image no. 2. 7.3 Sinus programs Posteroanterior Posteroanterior sinus exposure. Posteroanterior linear Posteroanterior linear sinus exposure. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 30 7 PROGRAMS Lateral Lateral exposure of the left or right sinus area. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 31: Preparations For Exposure

    - contact your service technician for help. If you have any reason to believe that the sensor might be faulty, take a test exposure before taking a patient exposure. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 32 This will secure the sensor in position. Push in the C-arm electrical connector button on the other side. This will make the electrical connection between the sensor and C-arm. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 33 For safety reasons, wait for at least ten seconds before you attach a sensor again. The blue indicator light under the locking knob has to switch off first. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 34: Preparing Planmeca Romexis

    8 PREPARATIONS FOR EXPOSURE 8.2 Preparing Planmeca Romexis First select the patient. Then select the panoramic exposure. Refer to the Planmeca Romexis User’s Manual for details on Romexis functions. 8.3 Preparing patient Ask the patient to remove any spectacles, hearing aids,...
  • Page 35: Panoramic Exposure

    NOTE Ensure that you insert the temple supports the right way round (i.e. with the convex side facing outward as shown). ProMax_2_3.eps • Use these patient supports. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 36 You can use these patient supports if the patient has a very large jaw. NOTE You can use these patient supports for edentulous patients or for patients who are unable to bite. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 37: Selecting Exposure Settings

    • left of the screen. Refer to section 7.1 “Panoramic programs” on page 22 for details. NOTE The available options depend on the X-ray unit model. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 38 NOTE The exposure values will automatically change according to the selected patient size. NOTE The patient size can also be selected on the next screen. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 39 It is not possible to expose two separate horizontal segments. NOTE The illustration on the screen is only an example. The actual size of the exposed area depends on the patient’s individual anatomy. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 40: Patient Positioning

    C-arm to move back. NOTE Planmeca ProMax 2D S2 X-ray units offer only one patient entry position. This entry position is the closed patient entry position where the C-arm is positioned around the temple supports.
  • Page 41 The positioning lights show on the patient’s face. Frankfort plane light Midsagittal plane light Layer light User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 42 To go to the next screen, select • this symbol • the forward button. Position the patient’s midsagittal plane so that it coincides with the midsagittal plane light. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 43 Check that the midsagittal plane light and the Frankfort plane light are still correctly positioned. Reposition them if necessary. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 44: Selecting Jaw Size And Shape

    (XS = 1, S & M & L = 2, XL = 3). NOTE If needed, you can change the patient size on this screen. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 45: Selecting Multiview (Planmeca Promax 3D X-Ray Units)

    9 PANORAMIC EXPOSURE 9.5 Selecting MultiView (Planmeca ProMax 3D X-ray units) NOTE MultiView is an optional feature for Planmeca ProMax 3D X-ray units. Use this button to select the MultiView imaging mode. This function allows you to adjust the viewing angle in the Planmeca Romexis program.
  • Page 46: Adjusting Exposure Values For Current Exposure

    Medium-sized adult (M) Large adult (L) Extra large adult (XL) MultiView SmartPan Child (XS) Small adult (S) Medium-sized adult (M) 12.5 Large adult (L) Extra large adult (XL) Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 47 Select the green check mark button. NOTE You can adjust the preset exposure values permanently as described in section 12.2.1 “Programs (2100)” on page 77. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 48: Selecting Dynamic Exposure Control (Dec) (Planmeca Promax 2D X-Ray Units)

    The setting can be adjusted between 20% (lower exposure values -> brighter image) and 200% (higher exposure values -> darker image). The recommended setting is 100% (default setting). Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 49: Taking An Exposure

    The image is shown on the computer screen. • Note that you must accept the image in the Planmeca Romexis program. Guide the patient away from the X-ray unit. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 50: Taking An Exposure With Autofocus (Planmeca Promax 2D S3 X-Ray Units)

    Make sure that you have selected the correct patient and exposure mode in the Planmeca Romexis program. NOTE Autofocus is an optional feature for Planmeca ProMax 2D S3 X-ray units. It is available for standard, interproximal and orthogonal panoramic programs.
  • Page 51 The image is shown on the computer screen. • Note that you must accept the image in the Planmeca Romexis program. Guide the patient away from the X-ray unit. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 52: Temporomandibular Joint (Tmj) Exposure

    Insert_temple_supports.eps NOTE Ensure that you insert the temple supports the right way round (i.e. with the convex side facing outward as shown). ProMax_2_3.eps Use these patient supports. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 53: Selecting Exposure Settings

    Refer to section 7.2 “Temporomandibular joint (TMJ) programs” on page 23 for details. NOTE The available options depend on the X-ray unit model. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 54 10.2.4 Selecting jaw side(s) Use this button to select the right, left or both jaw sides. 10.2.5 Selecting imaging angle (Planmeca ProMax 2D S3 & Planmeca ProMax 3D X-ray units) Use the arrow buttons to select the imaging angle for this exposure.
  • Page 55: Patient Positioning

    To switch them back on, do one of the following: Press the thumb wheel on the underside of the patient support table. Press any of the positioning controls (button or joystick). User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 56 10 TEMPOROMANDIBULAR JOINT (TMJ) EXPOSURE 10.3.3 Adjusting patient’s head position (Planmeca ProMax 2D S2 X-ray units) The positioning lights and the illustrations on the touch screen help you to position the patient’s head correctly. NOTE The illustrations are for guidelines only.
  • Page 57 Check that the midsagittal plane light and the Frankfort plane light are still correctly positioned. Reposition them if necessary. • FOR POSTEROANTERIOR EXPOSURES: It is not possible to adjust the imaging position. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 58 10 TEMPOROMANDIBULAR JOINT (TMJ) EXPOSURE 10.3.4 Adjusting patient’s head position (Planmeca ProMax 2D S3 & Planmeca ProMax 3D X-ray units) The positioning lights and the illustrations on the touch screen help you to position the patient’s head correctly. NOTE The illustrations are for guidelines only.
  • Page 59 (e.g. 1/2 = 77 mm and 2/2 = 67 mm). Check that the midsagittal plane light and the Frankfort plane light are still correctly positioned. Reposition them if necessary. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 60: Adjusting Exposure Values For Current Exposure

    Extra large adult (XL) You can adjust the preset exposure values (kV and mA) as described in section 9.6 “Adjusting exposure values for current exposure” on page 42. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 61: Taking An Exposure In Double Tmj Programs

    NOTE If needed, you can stop the imaging process by touching the red cross button after you have taken the first exposure. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 62 The image is shown on the computer screen. • Note that you must accept the image in the Planmeca Romexis program. Guide the patient away from the X-ray unit. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 63: Taking An Exposure In Three Angle Tmj Programs

    The image is shown on the computer screen. • Note that you must accept the image in the Planmeca Romexis program. Guide the patient away from the X-ray unit. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 64: Sinus Exposure

    Insert_temple_supports.eps NOTE Ensure that you insert the temple supports the right way round (i.e. with the convex side facing outward as shown). ProMax_2_3.eps Use these patient supports. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 65: Selecting Exposure Settings

    • left of the screen. Refer to section 7.3 “Sinus programs” on page 25 for details. NOTE The available options depend on the X-ray unit model. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 66: Patient Positioning

    Select the patient size as described in section 9.2.3 “Selecting patient size” on page 34. 11.2.4 Selecting jaw side for lateral exposures (Planmeca ProMax 2D S3 & Planmeca ProMax 3D X-ray units) Use this button to select the right or left jaw side. 11.3 Patient positioning 11.3.1 Selecting patient entry position...
  • Page 67 NOTE The illustrations are for guidelines only. To go to the next screen, select • this symbol • the forward button. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 68 30° X-ray unit with the height adjusting buttons. The patient’s back and neck must be straight. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 69: Adjusting Exposure Values For Current Exposure

    Extra large adult (XL) You can adjust the preset exposure values (kV and mA) as described in section 9.6 “Adjusting exposure values for current exposure” on page 42. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 70: Taking An Exposure

    The image is shown on the computer screen. • Note that you must accept the image in the Planmeca Romexis program. Guide the patient away from the X-ray unit. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 71: Settings

    Settings that can be entered by service personnel only (password required): • Technical To return to the main screen, select the settings symbol at the top left corner. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 72: User Settings

    Select User > 1200 Time and Date > 1210 Set System Time and Time / Date Display Format > Time Display Format. Select the display format you wish to use. Select the green check mark button. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 73 The time is set to the local time at the factory. Change the time setting to show the correct time before you start using the X-ray unit. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 74 Select the mode you wish to use. In demo mode you can practice or demonstrate the functions of the X-ray unit without radiation and PC connection. Select the green check mark button. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 75 Set the volume level to 0% if you do not wish to use this function. Select the green check mark button. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 76 C-arm to move back. NOTE Planmeca ProMax 2D S2 X-ray units offer only one patient entry position. This entry position is the closed patient entry position where the C-arm is positioned around the temple supports.
  • Page 77 X-ray system (indicator light on the hand-held exposure switch, wall exposure switch(es) and remote exposure lamp). Select the green check mark button. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 78 Select the network settings you wish to view. Select the green check mark button. NOTE Only a service technician or local administrator may change the network settings. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 79 Press and hold down the exposure button for the duration of the exposure. The C-arm will not move when you take a test exposure. Select the green check mark button. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 80 Select User > 1600 Clinic Management to view the network settings for the Planmeca Romexis Clinic Management module. NOTE Only a service technician or local administrator may change the settings. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 81: Program Settings

    Repeat for an other program type, patient size or image resolution (3D) if needed. Select the green check mark button. NOTE Always try to minimize the radiation dose to the patient. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 82 To manage program settings: Select Program > 2200 Program Features to manage program settings. For details on a specific setting, refer to the manual section that contains the corresponding function. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 83 (i.e. overrule your own settings in menu Programs (2100)). The preset exposure values are shown in sections “Adjusting exposure values for current exposure”. Select the green check mark button. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 84: About Tab

    • To view exposure statistics: Select About > 4200 Archive > Exposure Statistics to view statistical data about the X-ray unit. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 85 Note that when you enter the X-ray unit serial number, you have to include any letters shown at the beginning of the number. Select the green check mark button. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 86: Help Messages

    Check the slice count and / or slice 3. step between slices. H134 Patient safety area violation in tomo Check the slice count and / or slice 4. step between slices. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 87 No IP address defined for 3D sensor. H187 Problem during image data Exposure was interrupted. transmission. Contact your service technician for help. H189 The screen was touched during Exposure was interrupted. exposure. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 88: Error Messages

    Guide the patient away from the X-ray unit. Then clear the message by touching the green check mark. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 89: Cleaning And Disinfection

    Use a mild cleaning solution and a clean soft cloth for cleaning stains and dirt if needed. • Panoramic patient supports User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 90 15 CLEANING AND DISINFECTION • Cephalometric patient supports • Impression plate holder • Patient handles Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 91: Touch Screen

    15.2 Touch screen Wipe the touch screen regularly using a clean soft cloth that has been slightly dampened with water. NOTE Do not use any cleaning solutions. User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 92: Sensor

    NOTE FOR TOUCH SCREEN AND SENSOR: Refer to sections 15.2 “Touch screen” on page 87 and 15.3 “Sensor” on page 88. Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 93: Service

    16 SERVICE 16 SERVICE The X-ray unit must be checked by a qualified Planmeca service technician once a year or after every 10 000 exposures (if this is sooner). This will guarantee patient and user safety and ensure consistent image quality.
  • Page 94: Disposal

    - plastic PUR, other plastics Motors Component boards Cables, Copper, transformers steel, transformer oil X-ray tube Packing Wood, cardboard, paper, polystyrene Sensor Return sensor to Planmeca. Other parts Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 95: Technical Specifications

    Total 2.5 mm Al + 0.5 mm Cu SmartPan 3D s / 3D Classic / 3D Plus / 3D Mid: total 2.5 mm Al 3D Max: total 2.5 mm Al + 0.5 mm Cu User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 96 1 - 14 mA ±10% Pan / SmartPan 1 - 16 mA ±10% Scanning ceph 1 - 16 mA ±10% Planmeca ProCeph 16 mA ±10% mAs range Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 97 2D tomography: 1.5 3D Plus / 3D Mid: 1.4 SmartPan 3D s / 3D Classic: 1.27 3D Plus / 3D Mid / 3D Max: 1.4 Ceph 1.13 User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 98 3D Plus / 3D Mid 141 kg (311 lb) 3D Max 139 kg (306 lb) Scanning ceph 26 kg (57 lb) Planmeca ProCeph 20 kg (44 lb) Environmental requirements Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 99 3D Mid: 146 x 146 mm (5.74 x 5.74 in.) 3D Max: 193 x 242 mm (7.6 x 9.5 in.) SmartPan: Flat panel pixel size 127 m User’s Manual (2D Imaging) Planmeca ProMax 2D & 3D s & 3D Classic...
  • Page 100: Original Manufacturer

    No green objects next to X-ray 18.2 Original manufacturer PLANMECA Oy, Asentajankatu 6, FIN-00880 Helsinki, FINLAND Phone: +358 20 7795 500, Fax: +358 20 7795 555, www.planmeca.com Planmeca ProMax 2D & 3D s & 3D Classic User’s Manual (2D Imaging)
  • Page 102 Planmeca Oy | Asentajankatu 6 | 00880 Helsinki | Finland tel. +358 20 7795 500 | fax +358 20 7795 555 | sales@planmeca.com | www.planmeca.com...