5.2 Adjusting panoramic beam and patient positioning mechanism
Technical manual
The following sections list and describe the procedures for adjusting the
panoramic beam and patient positioning mechanisms.
There are two different adjustment procedures, depending on the X-ray unit
in use:
•
Recommended adjustment procedure using tomographic lights -
Intended for Planmeca ProMax X-ray units with software revision
3.8.0.0.R onwards
•
Traditional adjustment procedure using ruler - Optional for Planmeca
ProMax X-ray units with software revision 3.8.0.0.R onwards; mandatory
for Planmeca ProMax X- ray units with software revision 3.7.X.X and
lower