Technical manual
3. Touch Calibrations (3200) and select Lift Calibration (3230). The Lift
Calibration display appears.
4. Check the Sensor value field:
•
If the value in the field is green, the sensor is correctly calibrated.
•
If the value in the field is red, the sensor has to be calibrated, see the
next step.
7 Column adjustments
Planmeca ProMax 3D Max
183