System Conflicts (E8Xx) - Planmeca ProMax 3D Max with ProTouch Technical Manual

Hide thumbs Also See for ProMax 3D Max with ProTouch:
Table of Contents

Advertisement

3 Help and error messages
Error
Explanation
code
E711

3.2.6 System conflicts (E8xx)

Error
Explanation
code
E813
Missing internal parts
E814
E815
E816
E818
E819
E821
Software compatibility CPU FPGA version not compatible with
E822
E813
Missing internal parts
E814
E815
E816
E818
88
Planmeca ProMax 3D Max
3D Primary x-collimator opening is too
wide.
The following table lists the system conflict errors.
Missing internal parts, RTC not found.
Missing internal parts, CAM not found.
Missing internal parts, TUBE not found.
Patient position control keypad not
found.
Missing internal parts, Ethernet module
not found.
Ceph CA not found.
the main software.
CAM FPGA version not compatible with
main software.
Missing internal parts, RTC not found.
Missing internal parts, CAM not found.
Missing internal parts, TUBE not found.
Patient position control keypad not
found.
Missing internal parts, Ethernet module
not found.
Comments
Adjust the left blade
to the right or the right
blade to the left.
Comments
Check the
functionality of the
CAM PCB.
Check the cabling.
Check the
functionality of the
Ethernet PCB and
that it is attached
properly on the CPU.
Check the cabling
between the CEPH
PCB and CPU PCB.
Check the
functionality of the
CPU PCB.
Check the software
version of the ProMax
X-ray unit.
Check the
functionality of the
CAM PCB.
Check the cabling.
Check the
functionality of the
Ethernet PCB and
that it is attached
properly on the CPU.
Technical manual