Planmeca ProMax
®
3D s & 3D Classic with ProTouch
user's manual
3D imaging

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Planmeca ProMax 3D s and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Planmeca Planmeca ProMax 3D s

  • Page 1 Planmeca ProMax ® 3D s & 3D Classic with ProTouch user's manual 3D imaging...
  • Page 3: Table Of Contents

    11.2 Taking an exposure of an impression or plaster cast ..........47 SETTINGS ......................50 12.1 User settings ........................ 50 CLEANING .....................53 SERVICE ......................53 DISPOSAL .....................54 HELP MESSAGES ..................55 Planmeca ProMax 3D s & 3D Classic with ProTouch 1 User’s Manual (3D)
  • Page 4 IEC 60364 - equipment is used according to the operating instructions Planmeca pursues a policy of continual product development. Although every effort is made to produce up-to-date product documentation this publication should not be regarded as an infallible guide to current specifications. We reserve the right to make changes without prior notice.
  • Page 5: Introduction

    Make sure that you are fully acquainted with the appropriate radiation protection measures and these instructions before you use the X-ray unit. NOTE The X-ray unit may be used by health care professionals only. Planmeca ProMax 3D s & 3D Classic with ProTouch 1 User’s Manual (3D)
  • Page 6: Associated Documentation

    Planmeca Romexis program. The Romexis package contains the following manuals: • User’s Manual, Original English publication: 10014593 • Installation Manual, Original English publication: 10014600 2 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 7: Symbols On Product Labels

    Type B applied part (Standard IEC 60601-1) Separate collection for electrical and electronic equipment (Directive 2002/96/EC WEEE) Alternating current (Standard IEC 60417) Electrostatic sensitive device (Standard IEC 60417) Planmeca ProMax 3D s & 3D Classic with ProTouch 3 User’s Manual (3D)
  • Page 8: Safety Precautions

    NOTE When it is likely that evaluation of soft tissues will be required part patient’s radiological assessment, conventional CT or MR medical imaging should be used rather than CBCT. 4 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 9 Planmeca Romexis program. The last ten images can be manually imported into Romexis. NOTE Never place or hang any objects on any part of the X- ray unit. Planmeca ProMax 3D s & 3D Classic with ProTouch 5 User’s Manual (3D)
  • Page 10: Switching X-Ray Unit On

    On / off switch NOTE To prolong the lifetime of the X-ray unit, always switch the X-ray unit off when it is not in active use. 6 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 11: Main Parts

    MAIN PARTS MAIN PARTS General view of X-ray system Ethernet 1. X-ray unit 2. 3D reconstruction PC 3. Planmeca Romexis program Planmeca ProMax 3D s & 3D Classic with ProTouch 7 User’s Manual (3D)
  • Page 12: General View Of X-Ray Unit

    MAIN PARTS General view of X-ray unit C-arm 3D sensor Patient support table Patient handles Patient positioning controls Touch screen Telescopic column Stationary column Chair (included in delivery) 8 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 13: Sensors

    1. 3D sensor for Planmeca ProMax 3D s 2. 3D sensor for Planmeca ProMax 3D Classic 3. ProFace sensor for Planmeca ProMax 3D s and Planmeca ProMax 3D Classic Planmeca ProMax 3D s & 3D Classic with ProTouch 9 User’s Manual (3D)
  • Page 14: Patient Supports

    Head supports (A or B) 1. Adjustable head support 1. Head band 25 2. Temple pads for children 2. Support bars 3. Fastening straps 4. Support bars 10 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 15: Exposure Switch

    During exposure yellow radiation warning lights illuminate Ready on the exposure switch and on the touch screen. They 3) Yellow = Radiation indicate that the X-ray unit is generating radiation. Planmeca ProMax 3D s & 3D Classic with ProTouch 11 User’s Manual (3D)
  • Page 16: Emergency Stop Button

    A help message will appear on the touch screen. Guide the patient away from the X-ray unit. Then release the emergency stop button. The X-ray unit will automatically restart. 12 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 17: Touch Screen

    (or select an other option if available). NOTE Do not use sharp objects to operate the touch screen. Function selected Function not selected Planmeca ProMax 3D s & 3D Classic with ProTouch 13 User’s Manual (3D)
  • Page 18 NOTE You can switch demo mode on if you wish to practice or demonstrate the functions of the X-ray unit without radiation (Settings > User > 1300 Operational settings > 1310 User Mode > 1311 Set Demo Mode). 14 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 19: Patient Positioning Controls

    Open / close temple supports Press the temple support button to open the temple supports in 2D imaging. The temple supports can be closed by pressing the temple support button again. Planmeca ProMax 3D s & 3D Classic with ProTouch 15 User’s Manual (3D)
  • Page 20: Planmeca Promax 3D S Programs

    DOUBLE SCAN 2 x Ø50 x H50 mm 2 x Ø50 x H80 mm TRIPLE SCAN 3 x Ø50 x H50 mm 3 x Ø50 x H80 mm 16 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 21: Planmeca Promax 3D Classic Programs

    3 x Ø80 x H50 mm 3 x Ø80 x H80 mm 3D Models 8.2.1 Volume size Program Ø80 mm IMPRESSION Ø80 x H40 mm PLASTER CAST Ø80 x H40 mm Planmeca ProMax 3D s & 3D Classic with ProTouch 17 User’s Manual (3D)
  • Page 22: 3D Patient Exposure

    1. Push the sensor onto the connector on the C-arm. Sensor 2. Turn the locking knob over the fastening mechanism. This will secure the sensor in position. Locking knob 18 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 23 C-arm. C-arm electrical connector Sensor Locking knob 2. Turn the locking knob 180 degrees. This will release the locking mechanism. 3. Carefully pull the sensor out. Planmeca ProMax 3D s & 3D Classic with ProTouch 19 User’s Manual (3D)
  • Page 24 Head support A: Attaching adjustable head support Adjustable If you are using the adjustable head support, slide it onto head support the support bars. Support bars 20 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 25 50 mm (1.9 in.). Straps with a free length (i.e. when they are not stretched) of over 255 mm (10 in.) do not support the patient’s head firmly. Planmeca ProMax 3D s & 3D Classic with ProTouch 21 User’s Manual (3D)
  • Page 26 The lower the patient is positioned, the higher the image position will 22 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 27 -> 5 mm higher image position • Use the highest position and a chin cup when taking exposures of the teeth area. Teeth Tooth program program TEETH AREA: Chin cup up Planmeca ProMax 3D s & 3D Classic with ProTouch 23 User’s Manual (3D)
  • Page 28 Jossu 5 9.1.4 Preparing Planmeca Romexis First select the patient. Then click the 3D capture button. Refer to the Planmeca Romexis User’s Manual for details on Romexis functions. 24 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 29: Selecting Exposure Settings

    NOTE Selecting child patient (XS) will automatically reduce the volume size and patient dose. NOTE The exposure values will automatically change according to the selected patient size and image resolution. Planmeca ProMax 3D s & 3D Classic with ProTouch 25 User’s Manual (3D)
  • Page 30 The exposure values have been preset at the factory for each patient size and image resolution. The preset exposure values are average values and they are only meant to guide the user. 26 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 31 • To adjust the exposure values: Reduce Increase value value Adjust exposure values for this exposure Forward Fast forward (skip next screen) Planmeca ProMax 3D s & 3D Classic with ProTouch 27 User’s Manual (3D)
  • Page 32 NOTE In 3D Double Scan and 3D Triple Scan programs the selected area is the primary image volume. The other image volume(s) adjoin(s) the primary image volume. 28 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 33 Select jaw side (right / left / both sides) NOTE The available options depend on the selected program and image resolution. Planmeca ProMax 3D s & 3D Classic with ProTouch 29 User’s Manual (3D)
  • Page 34 X-ray image. Select the last button on the left of the screen to take both images at the same time. ProFace sensor Select 3D face photo 30 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 35: Preparing Patient

    (Settings > User > 1300 Operational settings > 1330 Patient positioning). This might be necessary if there is no space for the C-arm to move back. Planmeca ProMax 3D s & 3D Classic with ProTouch 31 User’s Manual (3D)
  • Page 36 • You can use fastening straps for additional head support if needed. Refer to section 9.1.2 “Attaching head support” on page 20 for details. 32 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 37: Adjusting Volume Position

    Press any of the positioning controls (button or joystick). Thumb wheel The lights will automatically switch off after two minutes. To switch them off earlier, press the positioning joystick. Positioning controls Positioning joystick Planmeca ProMax 3D s & 3D Classic with ProTouch 33 User’s Manual (3D)
  • Page 38 NOTE If the upper jaw half is selected, the lower edge of the image volume is positioned 30 mm above the volume bottom light. Upper jaw half selected 34 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 39 (i.e. the image volume center as seen from the front) will move accordingly. Left Right Right Left Image volume Positioning joystick (front light = X laser) Planmeca ProMax 3D s & 3D Classic with ProTouch 35 User’s Manual (3D)
  • Page 40 Patient support table Thumb wheel Front Back (side light = Y laser) Back Positioning joystick (side light = Y laser) Image volume Front 36 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 41: Taking A Scout Image Or 2D Views (Lat, Pa Or Lat-Pa)

    1. Select the view you wish to take. To take LAT-PA views, select both buttons (LAT and PA). The selected option is shown on the top of the forward button. Select scout or 2D view Planmeca ProMax 3D s & 3D Classic with ProTouch 37 User’s Manual (3D)
  • Page 42 Then take a new exposure as described above. Repeat the procedure until the image volume is in the correct place. NOTE Scout images are not saved in the Planmeca Romexis Lat-PA program. 38 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 43: Taking A 3D Exposure

    If you took several image volumes you must accept the image stitching function in the Planmeca Romexis program. Refer to the Romexis User’s Manual. 6. Guide the patient away from the X-ray unit. Planmeca ProMax 3D s & 3D Classic with ProTouch 39 User’s Manual (3D)
  • Page 44: 3D Face Photo

    1. Select the patient entry position as described in section 9.4.1 “Selecting patient entry position” on page 2. Guide the patient to the X-ray unit. The patient can sit or stand during the exposure. 40 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 45 NOTE When you adjust the light position the positioning light moves on the patient’s face. The positioning light on the illustration does not move according to your adjustment. Planmeca ProMax 3D s & 3D Classic with ProTouch 41 User’s Manual (3D)
  • Page 46: Taking A 3D Face Photo

    You hear a fast ticking sound when the photo is taken. 4. The photo is shown on the computer screen. 5. Guide the patient away from the X-ray unit. 42 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 47: 3D Model Exposure

    3. Let the material set. The setting time depends on the material used. Wait slightly longer than recommended in the instructions supplied by the manufacturer to ensure proper hardening. Planmeca ProMax 3D s & 3D Classic with ProTouch 43 User’s Manual (3D)
  • Page 48 (volume center lights, volume bottom light and incisor light) come on. The volume center Forward field lights cross in the middle of the image volume. 44 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 49 4. In the Planmeca Romexis program, click 3D > Model Capture. Refer to the Planmeca Romexis User’s Manual for details on Romexis functions. 5. Click the option Add Material in the window that appears. Planmeca ProMax 3D s & 3D Classic with ProTouch 45 User’s Manual (3D)
  • Page 50 OK. NOTE The calibration exposure values are automatically included at the beginning of the name. The exposure values shown here are only examples. 46 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 51: Taking An Exposure Of An Impression Or Plaster Cast

    (volume center lights, volume bottom light and incisor light) come on. The volume center Forward field lights cross in the middle of the image volume. Planmeca ProMax 3D s & 3D Classic with ProTouch 47 User’s Manual (3D)
  • Page 52 Then click the option Start Capture. NOTE Ensure that you have selected the correct exposure values on the X-ray unit. The exposure values shown here are only examples. 48 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 53 3D patient exposure. 3. The image is shown on the computer screen. NOTE The Romexis function Model Capture creates surface models (instead of voxel data images). Planmeca ProMax 3D s & 3D Classic with ProTouch 49 User’s Manual (3D)
  • Page 54: Settings

    To return to the main view, select the settings icon at the top left corner. 12.1 User settings 12.1.1 Language (1100) To change the language, select User > 1100 Language. Then select the language you wish to use. 50 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 55 Select User > 1200 Time and Date > 1210 Set System Time and Time / Date Display Format > Change System Time. Then use the plus and minus buttons to change the time. Planmeca ProMax 3D s & 3D Classic with ProTouch 51 User’s Manual (3D)
  • Page 56 Select User > 1200 Time and Date > 1220 Change System Date. Then select the day or use the arrow buttons to change the month or year. Next year Previous year Next month Previous month 52 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 57: Cleaning

    X-ray unit must be checked and recalibrated by a qualified Planmeca service technician once a year or after every 10 000 exposures if this is sooner. Planmeca ProMax 3D s & 3D Classic with ProTouch 53 User’s Manual (3D)
  • Page 58: Disposal

    - plastic PUR, other plastics Motors Component boards Cables, Copper, transformers steel, transformer oil X-ray tube Packing Wood, cardboard, paper, polystyrene Sensor Return sensor to Planmeca. Other parts 54 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 59: Help Messages

    Exposure was interrupted. exposure. Contact your service technician for help. H152 The line voltage is too low. Exposure is not possible. Contact your service technician for help. Planmeca ProMax 3D s & 3D Classic with ProTouch 55 User’s Manual (3D)
  • Page 60 H195 The 3D Model programs must not be Use the 3D Model programs for used for patient imaging. taking exposures of impressions or plaster casts only. 56 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 61: Error Messages

    Guide the patient away from the X-ray unit. Then clear the message by touching the green check mark. Planmeca ProMax 3D s & 3D Classic with ProTouch 57 User’s Manual (3D)
  • Page 62: Technical Specifications

    3D s / 3D Classic / 3D Mid: 1 - 14 mA ±10% 3D Max: 1 - 12.5 mA ±10% • Pan / SmartPan 1 - 16 mA ±10% 58 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 63 1700 mm (66.9 in.) Magnification: • 3D s / 3D Classic: 1.57 3D Plus / 3D Mid: 1.38, 1.44 or 1.80 3D Max: 1.38, 1.41 or 1.80 Planmeca ProMax 3D s & 3D Classic with ProTouch 59 User’s Manual (3D)
  • Page 64 Transport: • Temperature -20°C - +60°C • Relative humidity 10 - 90% RH (non-condensing) • Air pressure 700 - 1060 hPa Storage: • Temperature -10°C - +50°C 60 Planmeca ProMax 3D s & 3D Classic with ProTouch User’s Manual (3D)
  • Page 65 Frequency for fluorescent lamps 100 Hz • Even and uniform lighting • No natural light (no windows in the room) • No green objects next to X-ray Planmeca ProMax 3D s & 3D Classic with ProTouch 61 User’s Manual (3D)
  • Page 66: Original Manufacturer

    TECHNICAL SPECIFICATIONS 18.2 Original manufacturer PLANMECA Oy, Asentajankatu 6, FIN-00880 Helsinki, FINLAND Phone: +358 20 7795 500, Fax: +358 20 7795 555, www.planmeca.com 18.3 Dimensions ce_digi_etu_max2.eps Scanning ceph or Planmeca ProCeph 1145 mm 850 mm (45.1") (33.5") 270 mm (10.6")
  • Page 67: Minimum Operational Space Requirements

    59.1 in. 64.2 in. 61.4 - 93.9 in. X-ray unit with cephalostat 2150 mm 1630 mm 1560 - 2385 mm 84.6 in. 64.2 in. 61.4 - 93.9 in. Planmeca ProMax 3D s & 3D Classic with ProTouch 63 User’s Manual (3D)
  • Page 70 Planmeca Oy | Asentajankatu 6 | 00880 Helsinki | Finland tel. +358 20 7795 500 | fax +358 20 7795 555 | sales@planmeca.com | www.planmeca.com...