5.4.8 Panoramic mode patient positioning lights
Adjusting midsagittal plane light position
Adjusting layer light position
Adjusting Frankfort plane positioning light
5.4.9 Adjusting 3D lights position
Steps
Technical manual
Adjust the midsagittal plane light position as instructed in section "Adjusting
midsagittal plane light position" on page 142.
Adjust the layer light position as instructed in section "Adjusting layer light
position" on page 144.
Adjust the Frankfort plane positioning light as instructed in section "Adjusting
Frankfort plane light position" on page 147.
1. Enter the calibration mode.
1.a.
Touch the service spanner on the Main display.
1.b.
Touch Technical and enter password (1701).
1.c.
Touch Calibrations (3200) > Scara Arm (3250) > Scara Arm
Calibrations (3251).
2. Touch the Drive to pin point button.
3. Set the shoulder joint pin (1), and elbow and C-arm transportation pin (2)
to their position holes, as follows.
3.a.
Remove the shoulder arm upper cover, see section "Removing
shoulder arm upper cover" on page 199.
3.b.
Slide the transportation pin (1) through the hole in the shoulder
joint so that it goes into the positioning holes in the shoulder arm
and column.
5 Panoramic and 3D calibrations
Planmeca ProMax 3D Max
165