Calibration Errors (E7Xx) - Planmeca ProMax 3D Max with ProTouch Technical Manual

Hide thumbs Also See for ProMax 3D Max with ProTouch:
Table of Contents

Advertisement

Error
Explanation
code
E615
E616
E624
Network
(ETHERNET)
E627
E628
E629
E630

3.2.5 Calibration errors (E7xx)

Error
Explanation
code
E701
Missing calibration
E703
E704
E705
Tube filament
definition
E706
E710
Primary collimator
Technical manual
Error in communication with panoramic
sensor.
Error in communication with cephalostat
sensor.
Error in communication with external
GUI.
Error in communication with ProCeph.
Error in communication with 3D sensor.
Ethernet cable is disconnected or it is
broken.
Error in GUI - ProMax communication.
The following table lists the calibration errors.
Tube head filament definition not done.
Panoramic DEC not calibrated.
Cephalometric DEC not calibrated.
Tubehead filament calibration failure
(discontinuous tube current).
Tubehead filament calibration failure
(mA too low).
Primary y-collimator opening is too big.
3 Help and error messages
Comments
Check the sensor is
correctly installed to
the X-ray unit.
Check the sensor is
correctly installed to
the X-ray unit.
Check the network
connection and the
cabling.
Check the network
connection and the
cabling.
Check the Ethernet
cable connection and
cabling.
Reboot ProMax. If
problem persists,
check GUI - ProMax
cabling.
Comments
Perform tube filament
definition again.
Calibrate panoramic
DEC.
Calibrate
cephalometric DEC.
The current in the
tube head is not high
enough. Perform the
preheat calibration
again.
The current in the
tube head is not
raising on the voltage
area. Perform the
preheat calibration
again. Check the
functionality of the
tube head.
Adjust the lower blade
upwards or the upper
blade downwards.
Planmeca ProMax 3D Max
87