Planmeca ProMax 3D Max with ProTouch Technical Manual page 207

Hide thumbs Also See for ProMax 3D Max with ProTouch:
Table of Contents

Advertisement

To enable the removal of the column upper front panel
Technical manual
1. Enter the calibration mode:
1. Touch the service spanner on the Main display.
2. Touch Technical and enter password (1701).
2. Touch Calibrations (3200) and select Patient Support Calibration (3240).
The Patient Support calibration display appears.
8 Parts replacement and repair
Planmeca ProMax 3D Max
201