Download Print this page

Parameter Conversion - Keysight FieldFox B Series User Manual

Advertisement

NA (Network Analyzer) Mode
NA Mode Settings
All FieldFox settings and features are supported (except Parameter Conversion)
with mixed-mode S-parameters.
Learn more about Balanced Measurements with the FieldFox in the
FieldFox Supplemental Online Help:
https://www.keysight.com/us/en/lib/resources/service-manuals/keysight
-fieldfox-library-help-and-manuals-2153870.html
Library, Help and Manuals – FieldFox Online Supplemental Help).
How to make mixed-mode S-parameter measurements
1. Press Measure 1.
2. Then More.
3. Then choose from the following:
To make balanced reflection measurements at the DUT output, connect the
DUT output to the FieldFox ports.

Parameter Conversion

In NA Mode ONLY, converts the active S-parameter trace to an equivalent
impedance (Z), admittance (Y), or reciprocal 1/S-parameter
How to select parameter conversions
— Press Measure 1
— Then select an S-parameter
— Then
— Then
— Then choose from the following:
Keysight N9938-90006 User's Guide
Scc11
Common reflect/common incident for logical port 1.
Sdd11
Differential reflect/differential incident for logical port 1.
Sdc11
Differential reflect/common incident for logical port 1.
Scd11
Common reflect/differential incident for logical port 1.
More
Conversion
—Off
(default) No conversion is performed.
Z Conv
or
Y Conv
Perform conversion for impedance (Z) or
admittance (Y). Then choose from the following for either:
——Auto
The displayed S-parameter is converted to the
appropriate Z or Y parameter: Refl for S11 and S22; Trans for
S21 and S12. When the S-parameter is changed, the
appropriate conversion changes automatically.
(Keysight FieldFox
93

Hide quick links:

Advertisement

loading