SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)
SA Mode Settings
2. Enter a value from 1 to 10,000 using the numeric keypad, the
3. Press Enter

Alignments

To better reflect the enhancements implemented during the alignment
process, for firmware versions
referred to as "Channel Equalization" or "Channel Equalization Alignment"
(ChanEQ
This section contains the following:
Overview
The alignments softkeys are located by pressing the Cal 5 hardkey.
All Alignment or Align All Now corrects for receiver drift over temperature and
time. SA measurement applications monitor changes in internal temperature
and the time since last alignment update and trigger the need for a new
update when the alignment becomes stale. The alignment is deemed stale
after a significant change in temperature or a significant time elapsed from the
last update. The temperature change and time duration necessary to trigger an
alignment update is predetermined in the factory. Until the update is
performed, annotation indicating "questionable amplitude" is displayed
AMPTD?.
When All Alignment is enabled (by default) the FieldFox performs an
alignment process using the internal RF source. Although some sweeps are
delayed, measurement results are never disturbed.
If the Independent Source (tracking generator) is enabled for your
measurement, it will be borrowed to perform the alignment. Again, the
measurement results are not disturbed. Learn about the Independent Source in
"Independent Source/Tracking Generator " on page
The alignment process can be disabled. You may want to do this, for example,
if you are analyzing the amplitude stability of a signal.
SA and all modes built on its infrastructure require three types of alignments:
— InstAlign Amplitude Alignment
— RF Burst Amplitude Alignment
— Channel Equalization
204
or the rotary knob.
Alignment).
"Overview" on page 204
"All Alignment and Align All Now" on page 207
"Individual Alignments" on page 207
>
A.10.15 "IF Flatness Alignment" is now
184.
Keysight N9938-90006 User's Guide