Calibration for NA, CAT, and VVM Modes
1. Press
If an error appears ("Failure to compute calibration steps..."),
check to ensure that the frequency range of the Cal Kit covers the
frequency range of the measurement. You can verify the frequency range
of your Cal Kit at: www.keysight.com/find/fieldfoxsupport. Click Cal Kits.
2. Follow the Cal Wizard prompts. Connect the specified standard at the
3. At any time, press Back to re-measure a standard if you feel it was not
4. Press
CAL ON U
corrected with the Mechanical Cal.

ECal

ECals each have a recommended input power level. Depending on the
measurement setup, the FieldFox may default to an input power level that
exceeds your ECal's recommended input power level. Refer to the
reference guide for your ECal module to determine the correct input power
levels. Exceeding an ECal's recommended input power level results in a
compression and an invalid data condition. Refer to
www.keysight.com/find/ecal.
ECal is a complete solid-state calibration solution. Every ECal module contains
electronic standards that are automatically switched into position during a
calibration. These electronic standards have been measured at the factory and
the data stored within the memory of the ECal module. The FieldFox uses this
stored data, along with the measured data, to calculate the error terms for a
measurement calibration.
Keysight N9938-90006 User's Guide
Start
Calibration.
point where the DUT will be connected, then press Measure.
properly connected.
Finish
to complete the calibration.
is shown on the screen for all displayed measurements that are