Time Domain – Option 010
Gating
How to make Time Domain Gating settings
— Press Measure Setup 4
— Then
— Then
— Then choose from:
Start, Stop, Center, and Span Gate Times
These settings specify the time in the trace to be gated (in or out). The gate
times can be specified using either Start and Stop or Center and Span.
— Press
— A settings table appears that allows you to navigate and select the

Gating Type

This setting defines the type of filtering that will be performed for the gating
function. The gate start and stop flags on the display point toward the part of
the trace you want to keep.
Choose from the following:
Keysight N9938-90006 User's Guide
Trace 4 shows the frequency response with gating enabled.
Transform
Gating
On
Gating is being performed.
Off
Gating is NOT being performed.
Gating Start Stop
following settings. These settings are made exactly like those made on the
Transform Settings table. Learn how in
page
116.
Bandpass
KEEPS the responses within the Gating Start and Stop times.
Notch
REMOVES the responses within the Gating Start and Stop times.
"Transform Settings Table" on
123