SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)
Channel Measurements
How to select Measurement Preset:
— Press Preset >

Averaging

By default in ALL Channel measurements, averaging is enabled and set to
display the average of the last 15 measurements. When enabled, this average
setting is automatically making the following 'averaging' settings in order to
provide the most accurate power measurements:
— Averaging Type is set to Power.
— Trace State is set to Average.
— Detector is set to Average.
— Count is set to 15.
Any of these settings can be changed manually during a Channel
measurement. Learn more about these settings starting in
page
To change Averaging:
1. Press Meas Setup 4.
2. Then
3. Then
4. Enter a value between 1 and 10,000 using the numeric keypad,
5. Press Enter.

Traces

Only one measurement trace can be displayed in Channel Measurements.

Channel Power (CHP)

Channel Power measures total power over the specified Integrated BW. The
Integration Bandwidth (IBW) can be adjusted to measure the power over
multiple channels.
218
Mode Preset
202.
Average Enable ON OFF
Average Count
arrows, or the rotary knob.
OFF effectively sets the Average Count to 1.
Keysight N9938-90006 User's Guide
"Average Type" on